期刊文献+

用于医学超声内窥成像的编码激励技术 被引量:1

Coded Excitation for Endoscopic Ultrasonography
下载PDF
导出
摘要 为了提高医学超声内窥系统中图像的信噪比,将编码激励技术引入超声成像系统.仿真研究了编码长度和换能器相对带宽对成像信噪比(SNR)提升的影响.发现在采用长度较短的2-5位Barker编码激励时,成像信噪比的提升在编码长度为4位时达到峰值,且信噪比的提升随换能器相对带宽的增加而增加.基于超声内窥成像系统的编码激励实验表明,在峰值激励电压为25 V和换能器相对带宽为20%的情况下,采用编码激励技术能够获得1.85 dB的成像信噪比提升. In order to enhance the signal-to-noise ratio ( SNR), coded excitation has been implemented for endoscopic ultrasonography. With Barker code excitation, the influences of the code length and the fractional bandwidth of the transducer on the enhancement of SNR have been carefully investigated by means of numerical simulation. For relatively short Barker code with bit length of 2--5, the SNR gain increased with the increase of the transduce fractional bandwidth reached its maximum when the bit length was 4. An experiment was conducted to verify the efficiency of coded excitation. For Barker code with bit length of 4, the SNR gain was 1.85 dB when excitation voltage was 25 V and transducer fraction band- width was 20%.
出处 《纳米技术与精密工程》 EI CAS CSCD 2009年第6期505-509,共5页 Nanotechnology and Precision Engineering
基金 教育部重点基金资助项目(106048) 国家高技术研究发展计划(863)项目(2007AA04Z339) 国家自然科学基金资助项目(60827005)
关键词 编码激励 超声成像 Barker码 信噪比 换能器 coded excitation ultrasound imaging Barker code signal-to-noise ratio(SNR) transducer
  • 相关文献

参考文献10

  • 1肖哲,陈晓冬,李婉晖,郁道银.医用内窥镜超声探头驱动系统的设计[J].传感技术学报,2005,18(3):642-645. 被引量:17
  • 2冯若主编..超声诊断设备原理与设计[M].北京:中国医药科技出版社,1993:1251.
  • 3宋光德,刘娟,栗大超.电容式微加工超声传感器(cMUT)的结构设计及仿真[J].纳米技术与精密工程,2005,3(2):156-159. 被引量:4
  • 4Wells P N T. Current status and future technical advances of ultrasonic imaging [ J ]. IEEE Engineering in Medicine and Biology, 2000, 19(5): 14-20. 被引量:1
  • 5Chiao R Y, Hao X H. Coded excitation for diagnostic ultrasound : A system developer's perspective[ J]. IEEE Transactions on Ultrasonics, Ferroelectrics , and Frequency Control, 2005, 52(2) : 150-170. 被引量:1
  • 6刘凯,高上凯.高频超声成像中的编码激励研究[J].北京生物医学工程,2007,26(1):31-35. 被引量:10
  • 7Carlson A B, Crilly P B, Rutledge J C. Communication Systems: An Introduction to Signal and Noise in Electrical Communication[ M ]. Beijing : Higher Education Press ,2002. 被引量:1
  • 8Welch L R, Fox M D. Practical spread spectrum pulse compression for ultrasonic tissue imaging [ J ]. IEEE Transactions on Ultrasonics, Ferroeleetrics , and Frequency Control, 1998, 45 (2) : 349-355. 被引量:1
  • 9Morimichi Fukuda, Kenichiro Hitata, Hiroshi Natori. Endoscopic uhrasonography of the esophagus [ J ]. World Journal of Surgery, 2000, 24(2) : 216-226. 被引量:1
  • 10鲍静..医学超声成像系统的编码激励技术研究[D].天津大学,2007:

二级参考文献10

  • 1[1]Jin Xuecheng, Ladabaum Igal, Degertekin F Levent, et al.Fabrication and characterization of surface micromachined capacitive ultrasonic immersion transducers [ J ]. Journal of Microelectromechanical Systems, 1999,8 ( 1 ): 100-114. 被引量:1
  • 2[2]Ladabaum I, Jin X, Soh H T, et al. Microfabricated ultrasonic transducers: Towards robust models and immersion devices [ A ]. In: Ultrasonics Symposium [ C ]. San Antonio,USA, 1996. 335-338. 被引量:1
  • 3会长善.超声工程[M].哈尔滨工业大学出版社,1989.109-111. 被引量:2
  • 4Newhouse VL,Bendick PJ.An ultrasonic random signal flow measurement system.Jour Acoust Soc of America,1974,56:860-870. 被引量:1
  • 5Matthew O' Donnell.Coded Excitation System for Improving the Penetration of Real-Time Phased-Array Imaging Systems.IEEE Transactions on Ultrasonics,Ferroelectrics,and Frequency Control,1992,39(3):341 -351. 被引量:1
  • 6Welch LR,Fox MD.Practical spread spectrum pulse compression for ultrasonic tissue imaging.IEEE Transactions on Ultrasonics,Ferroelectrics,and Frequency Control,1998,45:349-355. 被引量:1
  • 7Yang Mo Yoo,Woo-Yul Lee,Tai Kyong Song.A low voltage portable system using modified Golay sequences.Ultrasonics Symposium,IEEE,2001,2:7-10. 被引量:1
  • 8Ruprecht J,Rupf M.On the search for good aperiodic binary invertible sequences.IEEE Trans on Inform Theory,1996,42:1604-1612. 被引量:1
  • 9Richard Y,Chiao,Xiaohui Hao.Coded excitation for diagnostic ultrasound:A system developer 's perspective.IEEE Transactions on Ultrasonics,Ferroelectrics,and Frequency Control,2005,52(2):160-170. 被引量:1
  • 10孙军华,董明利,祝连庆.钢轨高速探伤系统超声发射/接收装置的设计[J].工具技术,2002,36(12):51-53. 被引量:8