摘要
介绍一套“有源法”光时域反射计(OTDR)衰减特性检定装置。对参考量值采取自标定的方法,用一种串联替代法实现对高重复性衰减器3.010dB值的准确标定。该装置检定灵活性强,对OTDR衰减特性检定的不确定度优于0.02dB/dB(95%置信度)。
狝n OTDR attenuation calibration apparatus is developed. Used as a reference, the 3.010dB attenuation of a high repeatability optical attenuator is calibrated using a self-calibration method in a serial substitute way. The apparatus is flexible for calibrating OTDR, the uncertainty attained is better than 0.02dB/dB (95% confidence level).
出处
《计量学报》
CSCD
北大核心
1998年第3期203-206,共4页
Acta Metrologica Sinica