期刊文献+

氢化物发生—原子荧光光谱法测定工业磷酸中的铅 被引量:3

Determination of Lead in Phosphoric acid for Industry useby hydride generation-atomic florescence spectrophotometry
下载PDF
导出
摘要 采用盐酸处理工业磷酸试样,用氢化物发生-原子荧光光谱法测定工业磷酸中的铅,对仪器条件、盐酸酸度、预还原剂用量、还原剂用量、共存元素干扰进行了试验,方法的检出限为0.028ug/L,测定精密度0.84%,回收率为98.9%-101.4%。建立的氢化物发生-原子荧光光谱法测定工业磷酸中的铅的分析方法具有操作简单,灵敏度高,重现性好等优点,能满足日常检验的要求。 The content of Lead in phosphoric acid for industry use was determined by hydride generation - atomic florescence spectrophotometry ( HG - AFS). The samples were treated by hydrochloric acid. The instrument conditions and influence of the concentration of hydrochloric acid, potassium borohydride and coexisting elements were investigated. The detection limit of the method was 0.028ug/L with relative standard deviation of 0.84% and recovery of 98.9 - 101.4%. The method has high sensitivity,its precision and accuracy were both meet requirements, can be used in routine inspection of Lead in Phosphoric acid for industry use.
机构地区 钦州检验检疫局
出处 《贵州化工》 2009年第2期42-44,53,共4页 Guizhou Chemical Industry
关键词 氢化物发生-原子荧光光谱法 工业磷酸 Hydride generation -atomic florescence spectrophotometry Phosphoric acid for industry use Lead
  • 相关文献

参考文献10

二级参考文献40

共引文献89

同被引文献22

  • 1孙景章.原子吸收光谱法测定污水中铅的最佳操作条件研究[J].环境工程,2004,22(4):53-55. 被引量:3
  • 2龙萍,梁骏.湿法磷酸中钾、钠、钙、镁的原子吸收光谱法测定[J].化学研究与应用,2006,18(7):837-839. 被引量:8
  • 3SN/T2049-2008.进出口食品级磷酸中铜、镍、铅、锰、镉、钛的测定电感耦合等离子体原子发射光谱法[S]. 被引量:3
  • 4夏玉宇.化验员实用手册[M].北京:化学工业出版社,2005.515. 被引量:22
  • 5SO 6678--1981 Phosphoric acid for industrial use Determination of lead content Atomic absorption spectrometric method [S]. 被引量:1
  • 6Sonia D, Nicola V, Costanza M, et al. Dynamic re- action cell ICP-MS for determination of total As, Cr, Se and V in complex matrices.. Still a challenge? A review [J]. Anal. Chim. Acta. ,2011,698(1-2), 6-13. 被引量:1
  • 7Denis P, Matthias L, Jiirgen W. Reduction of poly atomic interferences in biological material using dy namic reaction cell ICP-MS [J] Microchem. J. 2010,95(2), 315-319. 被引量:1
  • 8Ali K, Laurent N, Rachida C, et al. Optimization of ICP-MS collision/reaction cell conditions for the de- termination of elements likely to be interfered (V, Cr, Fe, Co, Ni, As and Se) in foodstuffs[J]. Ta- lanta,2011, 85(5), 2605-2613. 被引量:1
  • 9Rodrigo F S S, Marcelo B B G, Edenir R P, et al. Performance evaluation of collision-reaction interface and internal standardization in quadrupole ICP-MS measurements [J]. Talanta,2011, 86, 241-247. 被引量:1
  • 10Guo W, Hu S, Xiao Y, Zhang H, et al. Direct de- termination of trace cadmium in environmental sam- ples by dynamic reaction cell inductively coupled plasma mass spectrometry [J]. Chemosphere., 2010, 81(11), 1463-1468. 被引量:1