摘要
目的 评价应用国产偏心型室间隔缺损封堵器经导管治疗嵴内型 (IVSD)和肺动脉瓣下型室间隔缺损 (SVSD)的疗效和安全性。方法 2 8例室间隔缺损患者 ,其中IVSD 2 2例 ,肺动脉瓣下型缺损6例。男 15例 ,女 13例 ,年龄 3~ 34岁 ,平均 (12 .4± 5 .9)岁。封堵器由两个圆形盘片和连接腰组成 ,腰部长 2mm ,两侧盘片呈不对称型 ,左心室侧的盘片呈偏心型 ,靠主动脉侧的边缘为 0mm ,与其相对的边缘为 6mm ,右心室侧的盘片比腰部直径大 4mm。在X线透视和经胸超声引导下 ,经 7~ 10F鞘管通过股静脉途径放置封堵器。结果 左心室造影测量的缺损直径 2~ 12mm ,平均 (4 .5 4± 1.93)mm。缺损上缘距主动脉瓣 0~ 1mm ,平均 (0 .35± 0 .4 5 )mm。 2 2例IVSD患者封堵全部成功 ,6例SVSD中 4例封堵治疗成功 ,1例合并微量主动脉瓣返流。 2例因缺损大而放弃封堵治疗。未发生其他并发症。X线暴露时间9~ 4 0min ,平均 (14 .0 7± 5 .19)min ,操作时间 4 0~ 15 0min ,平均 (5 9.81± 17.76 )min。结论 国产偏心型室间隔缺损封堵器可以成功封堵IVSD和部分SVSD ,近期疗效佳 。
Objective To assess the efficacy and safety of transcatheter closure of intracristal ventricular septal defect (IVSD) and subpulmonary ventricular septal defect(SVSD)with home-made eccentric nitinol occluder.Methods Transcatheter closure was attempted in 28 patients (15 men and 13 women), with IVSD(n=22)and SVSD(n=6) at a age of 12.44±5.86 years (ranged from 3 to 34 years of age) The homemade device consists of two low profile disks made of Nitinol wire mesh with a 2 mm connecting waist. The left disk is 6 mm towards the apex and 0 mm towards the aortic valve. The right disk is 4 mm larger than the waist. The devices were deployed via the femoral vein using 7-10 Fr sheaths with the guidance of transthoracic echocardiography and fluoroscopy. Results The VSD diameter was 4.54 ±1.93 mm (ranged 2 to 12 mm). the distance of VSD to aortic valve was 0.35±045 mm (range 0 to 1 mm )。The connecting waist diameter of device was 7.65±3.11 mm(ranged 4 to 14 mm). The device was implanted successfully in 26 of 28 procedures. There was a trivial aortic regurgtation in 1 patient with SVSD after deployment of the occluder. Implantation was unsuccessful in 2 patients with SVSD, with having large defects. No other complications were observed. The mean fluoroscopy time for the procedure was 14.07±5.19 min (range 9 to 40 min). The procedure time was 59.81±17.76 min (range 40 to 150 min).Conclusions Transcatheter closure of IVSD and SVSD with homemade eccentric nitinol occluder is effective, easy and safe.Longer follow-up is required to assess long-term efficacy.
出处
《介入放射学杂志》
CSCD
2004年第S2期139-140,共2页
Journal of Interventional Radiology
关键词
经导管
嵴内型室间隔缺损
封堵器
Ventricular septal defect
Catheterization
Occluder