期刊文献+

^(239)Pu α 粒子的微剂量谱和径向剂量分布

MICRODOSIMETRIC SPECTRA AND RADIAL DOSE DISTRIBUTION OF 239 Pu α PARTICLES
下载PDF
导出
摘要 用大容积(60mm×80mm)的圆柱型流气式组织等效正比计数器测定了239Puα粒子的微剂量谱和径向剂量分布。α粒子束在无窗的情况下直接穿入正比计数器,由离子穿过而发生能量沉积的机率Rb为1。测量结果证实:沿α粒子径迹由单次事件沉积的剂量平均比能z1D(Gy)与距径迹核径向距离b(μm)之间完全符合z1D=ab-2的函数关系,同时给出了求出最大径向距离(penumbra)的方法。 Microdsimetric spectra and radial dose distribution of 239 Pu α particles were measured by the cylindrical flowing tissue equivalent proportional counter with larger volume. Because of the best collimated beam traverse in TEPC (tissue equivalent preportional courreer), the probability of energy deposition per ion traversal R b is 1. The measured dose mean specific energy z 1D from single event varies strictly in accordance with formula z 1D = ab -2 , b is the radial distance from track core. The method of determination of the coefficient a is given.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 1998年第2期168-172,共5页 Atomic Energy Science and Technology
基金 国家自然科学基金
关键词 正比计数器 微剂量谱 径向剂量分布 钚239 Tissue equivalent proportional counter Microdosimetric spectrum Radial dose distribution Maximum radial distance
  • 相关文献

参考文献4

二级参考文献3

  • 1田志恒,辐射剂量学,1992年,68页 被引量:1
  • 2李德平,辐射防护手册.第3分册,1990年,9页 被引量:1
  • 3团体著者,原子核物理实验方法.上,1981年 被引量:1

相关作者