摘要
目的:总结应用自体心包夹涤纶条的方法治疗重度三尖瓣关闭不全的经验。方法:2001-07至2007-10间67例重度三尖瓣关闭不全的患者术中全部应用自体心包夹涤纶条间断缝合将其固定于三尖瓣环上,根据瓣环扩大部位及程度决定该处缝缩多少。结果:手术后早期死亡1例。44例术后无三尖瓣反流或仅有极少量反流,16例术后轻度反流,7例为中度反流。随访3~72个月,平均46.5个月,随访期间,术后无反流的患者中有2例患者出现轻度反流,术后轻度反流的患者8例未出现反流,术后中度反流的患者4例变为轻度反流。1例患者术后3年死于脑出血。结论:自体心包夹涤纶条三尖瓣成型术是纠正重度三尖瓣关闭不全的有效方法。
Objective: To summarize the experience for treating severe incompetent tricuspid valve by Dacron strip covered with autolo- gous perieardimn.
Methods : Sixty seven patients admitted to our hospital from July 2001 to October 2007 were enrolled in this study. The Dacron strips covered with autologous pericardium were anchored to the tricuspid annulus by interrupted stitches to repair severe incompetent tricuspid valves.
Results : Among 67 patients, 44 cases had not or had only mild tricuspid regurgitation after the operation, 7 had moderate regurgitation after the operation, and 16 had mild incompetent tricuspid valve. All patients were followed up for 3 -72 months in the average of 46. 5 months. 2 patients without regurgitation at peri-operation period developed mild regurgitation during the follow-up time. 8 patients with mild regurgitation during peri-operation showed no regurgitation at the follow up time. 4 patients with moderate regurgitation during peri-operation changed to mild regurgitation at follow-up time. For all those 67 patients, 1 died after the operation and another died 3 years after operation with cerebral hemorrhage.
Conclusion: Tricuspid valve annuloplasty by Dacron strip covered with autologous pericardium was an effective method for treating severe incompetent tricuspid valve.
出处
《中国循环杂志》
CSCD
北大核心
2008年第6期456-458,共3页
Chinese Circulation Journal