消化道诊疗胶囊三维磁定位系统的误差修正

Error correction of three dimensional magnetic localization system used for tracking micro diagnosis capsule in gastrointestinal tract
下载PDF
导出
摘要 设计了基于三轴磁感传感器阵列的体内微型诊疗胶囊定位跟踪系统,分析了系统的主要误差来源,提出了对背景磁场干扰、传感器灵敏度差异、传感器位置偏差、三轴传感器模块的非理想正交等误差的补偿和修正方法,并进行了定位实验和轨迹跟踪实验以检验误差修正效果。实验结果表明:误差修正后系统的定位与定向精度得到显著提高,对于距离传感器阵列平面150 mm的φ8 mm×8 mm的圆柱形N35铷铁硼永磁体,平均定位与精度达到8mm以内,平均定向精度达到6°以内,并能够对沿模拟肠道运动的内置永磁体的电子胶囊进行轨迹跟踪。 A 3-axis magnetic inductive sensor array system used for localization and tracking of micro diagnosis capsule in human gastrointestinal tract is designed, error sources of this tacking system are analyzed, and the error correction methods are proposed to obtain better accuracy. The correction methods include background magnetic field elimination, sensitivity calibration, sensor position correction and sensor orthonormality adjustment. The experiment results show that the error correction procedure can improve the localization accuracy prominently. For the Ф8 mm ×8 mm cylindrical N35 NdFeB magnet,the average position error is less than 8 ram, and orientation error is less than 6°when the magnet is 150 mm above the sensor array plane after the error correction procedure is implemented. The localization system can also track the locus of the capsule with a magnet inside, when it moves among the simulative intestinal tract.
出处 《传感器与微系统》 CSCD 北大核心 2008年第11期93-96,99,共5页 Transducer and Microsystem Technologies
基金 国家自然科学基金资助项目(30700160) 重庆市自然科学基金资助项目(CSTC2006BA5005)
关键词 磁定位 磁测量 磁传感器 误差修正 微型诊疗装置 magnetic localization magnetic measurement magnetic sensor error correction micro diagnosis and treatment devices
  • 相关文献

参考文献14

二级参考文献24

  • 1丁振,潘泉,吴锦涛,张洪才.最小二乘被动式跟踪算法与仿真[J].数据采集与处理,1996,11(2):127-130. 被引量:1
  • 2孟桥,陆佶人.综合二重最小二乘被动定位算法[J].电子学报,1997,25(5):106-109. 被引量:1
  • 3张玉民 戚伯云 等.电磁学[M].北京:中国科学技术大学出版社,1997.25. 被引量:6
  • 4邱立军.[D].武汉:海军工程大学,2000:11~12. 被引量:2
  • 5Hine. A Magnetic Compasses and Magnetometers[M]. London:Adam Hilger LTP, 1968. 270. 被引量:1
  • 6W Weitschies,J Wedemeyer,et al. Magnetic markers as a noninvasive tool to monitor gastrointestinal transit.IEEE Transaction on Biomedical Engineering, 1994,41(2) :192-195. 被引量:1
  • 7Casper R,et al. Medical capsule device activated by radio-frequency(RF) signal. US Patent 5170801. 被引量:1
  • 8Haynor,et al. System and method to determine the location and orientation of an mdwelling medical device. US Patent 5879297. 被引量:1
  • 9Golden,et al. Apparatus and method for locating a medical tube in the body of a patient. US Patent 5622169. 被引量:1
  • 10林春生.[D].武汉:海军工程大学,1996:59~60. 被引量:6

共引文献158

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部