期刊文献+

用RP-HPLC法测定人体内辛伐他汀血药浓度 被引量:3

Determination of human plasma concentration of simvastatin by RP-HPLC method
下载PDF
导出
摘要 目的:建立RP-HPLC法测定人血浆中辛伐他汀的含量。方法:采用Kromasil C18色谱柱(250mm×4.6mm,5μm);流动相为甲醇-乙腈-水(80:5:15);流速为1.0mL/min;检测波长238nm;柱温为30℃。结果:辛伐他汀的血药浓度在2.0~150.0ng/mL范围内线性良好,定量下限为2.0ng/mL。回归方程为A=1.589×10^8c-218.0,r=0.9999(n=7),低、中、高浓度(6.5、23.0、65.0ng/mL)的相对回收率分别为(100.28±3.35)%、(97.57±3.21)%和(100.25±2.82)%,日内和日间RSD为2.40%~4.32%(n=5)。结论:本法快速、高效、灵敏,适用于辛伐他汀的血药浓度测定。 Objective: To establish a RP-HPLC method for determination of concentration of simvastatin in human plasma. Methods:Chromatography was carried out on a Kromasil C18 column, using methanol-aeetonitrile-water (80 : 5 : 15) as mobile phase at a flow rate of 1.0 mL/min. The detection wavelength was set at 238 nm. Column temperature was kept at 30 ℃ Results: The linearity range of simvastatin was 2.0-150.0 ng/mL and the quantitation limit was 2.0 ng/mL. The regressive equation was A=1. 589 × 10^3 c-218.0, r=0. 999 9 (n = 7). The relative recovery rates of low, middle and high concentrations (6.5,23.0,65.0 ng/mL) were (100.28±3.35)%,(97.57±3.21)% and (100.25±2.82)%,respectively. Intra- and inter-day RSD was within 2.40%-4. 32% (n = 5). Conclusion: The method is rapid, efficient, sensitive and suitable for determination of simvastatin in human plasma.
出处 《药学服务与研究》 CAS CSCD 2008年第5期370-372,共3页 Pharmaceutical Care and Research
关键词 色谱法 高压液相 血药浓度 辛伐他汀 chromatography,high pressure liquid plasma concentration simvastatin
  • 相关文献

参考文献7

二级参考文献11

共引文献614

同被引文献29

引证文献3

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部