摘要
研究了多元校正方法——偏最小二乘(PLS)在波长色散X射线荧光光谱分析中的应用,并将该法与传统的经验系数法(ECM)进行了比较.其考察对象为一组转炉渣文献数据、一组不锈钢样和一组锌精矿样.结果表明,PLS比ECM准确、稳定.经分析探讨,认为其主要原因是PLS能滤除噪音.文中建立了描述浓度和强度本质的关系式.
In the paper,a kind of multivariate calibration method,the partial least squares regression(PLS),used for the wavelength dispersive X ray fluorescence(XRF) spectrometry is studied,and compared with the traditional empirical coefficient methods(ECM).These studies relate to the intensities and concentrations of converter slags from literature,a set of stainless steel samples,and a set of zinc concentrate samples.The results show that the PLS is more accurate and more stable than ECM.The major reason is that the PLS calibration can filter noise,and set up the intrinsical relations between the concentrations and their intensities.
出处
《中南工业大学学报》
CSCD
北大核心
1997年第6期595-598,共4页
Journal of Central South University of Technology(Natural Science)
关键词
X射线荧光光谱
偏最小二乘
数学校正
稳定性
X ray fluorescence spectrometry
partial least squares regression
multielement analysis
mathematical correction