X线摄影焦点(F)-IP板的距离对CR成像质量的影响 被引量:3

Influence of Focus(F)-IP Plate Range in X-ray Photography on CR Image Quality
下载PDF
导出
摘要 目的探索X线管焦点(F)-IP板的最佳成像距离,提高CR图像质量。方法根据X线管焦点放大率和几何成像原理,求出最小焦点(F)-IP板的距离;综合考虑X机性能、被照体厚度和不同摄影部位,制定各部位摄影距离。结果现摄影距离CR图像质量明显提高,特别是四肢摄影X线曝光量显著降低。结论不同被照体厚度或不同部位,采用不同的摄影距离,在明显提高CR图像质量的同时,显著降低X线曝光量,减少被检者受照剂量,降低X线机输出功率,延长X线机使用寿命。 Objective To evaluate into the best range for imaging in X-ray focus (F)-IP plate and to improve CR image quality. Methods According to X-ray focus magnifying power and geometrical imaging principal,the range of minimal focus (F)-IP plate was obtained.Each CR photograph range was determined, combined with X-ray machine function, objective thickness and various position. Results Photograph range for CR image was highly improved, esp. X-ray exposure of limbs photography was reduced significantly. Conclusion As the objective thickness or location is different, the photography range is different. When CR image quality is improved, X-ray exposure is reduced. Service time of X-ray system is prolonged.
出处 《中国医疗设备》 2008年第9期137-138,共2页 China Medical Devices
关键词 摄影距离 图像质量 X线摄影 photograph range image quality radiography
  • 相关文献

参考文献6

  • 1祁吉,(日)高野正雄主编..计算机X线摄影[M].北京:人民卫生出版社,1997:99.
  • 2官秋青,赵晓红,李清军.曝光条件与辐射剂量[J].医学影像学杂志,2004,14(9):751-752. 被引量:5
  • 3熊维军,杨宏美,夏兆云.影响CR影像质量的因素与对策[J].武警医学,2005,16(12):932-934. 被引量:3
  • 4袁聿德主编..X线摄影学[M].北京:人民卫生出版社,1993:245.
  • 5王溱主编,于荣溥等编写..X线诊断学[M].石家庄:河北教育出版社,1994:945.
  • 6孙秀兰等主编..放射医学与防护[M].济南:济南出版社,2001:360.

二级参考文献4

共引文献6

同被引文献25

  • 1孔祥闯,夏忠心,余建明.婴幼儿胸部DR摄影距离对体表入射剂量与图像质量的影响[J].中华放射医学与防护杂志,2011,31(6). 被引量:9
  • 2吴长春,范朝梅.缩短焦片距增加放射防护效果[J].中华放射医学与防护杂志,2006,26(2):185-185. 被引量:4
  • 3赫英辉,邓振生,高凯,余厚军,张学昕,贺洪德,牟轩沁,孙立军.附加铝质滤过板对X线图像质量和患者吸收剂量的影响[J].第四军医大学学报,2006,27(11):1002-1004. 被引量:12
  • 4国家质量监督检验检疫总局.GB18871-2002电离辐射防护与辐射源安全基本标准[S].北京:中国标准出版社,2002. 被引量:50
  • 5Robinson J, McLean D. Extended focal-film distance technique : an analysis of the factors in dose reduction for the AP knee radiograph [ J ]. Radiography,2001,7 (3) : 165 - 170. 被引量:1
  • 6Dilger R, Egan I, Hayek R. Effects of Focus Film Distance (FFD)variation on entrance testicular dose in lumbar-pelvic radiography [ J ]. Australas Chiropr Osteopathy, 1997,6 ( 1 ) : 18 - 23. 被引量:1
  • 7Brennan PC, Nash M. Increasing FFD: an effective dose-reducing tool for lateral lumbar spine investigations[ J]. Radiography, 1998,4 (98) :251 -259. 被引量:1
  • 8Mourik JE, Vander Tol P, Veldkamp WJ, et al. Comparison of wireless detectors for digital radiography systems : image quality and dose[ J ]. Radiat Prot Dosim, 2016, 169 ( 1-4 ) : 303-307. DOI: 10. 1093/xtd/ncv450. 被引量:1
  • 9Brennan PC, McDonnell S, O' Leary D. Increasing film-focus distance(FFD) reduces radiation dose for X-ray examinations[ Jl - Radiat Prot Dosim, 2004, 108 ( 3 ) : 263-268. DOI: 10. 1093/ rpd/nch029. 被引量:1
  • 10Robinson J, Mclean D, Extended focal-film distance technique: an analysis of the factors in dose reduction for the AP knee radiography [ J ]. Radiography, 2001, 7 ( 3 ) : 165-170. DOI: 10. 1053/radi. 2001. 0324. 被引量:1

引证文献3

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部