期刊文献+

挥发性有机污染物实时在线化学电离质谱测量方法研究 被引量:3

Study of On-line Organic Pollutants Measurement with Chemical Ionization Mass Spectrometry
下载PDF
导出
摘要 以《大气污染物综合排放标准))(GB 16297-1996)中有机物的实时在线测量为目的,对自行研制的质子转移反应质谱(proton transfer reaction mass spectrometry,PTR-MS)装置的参数进行了优化和测量。利用高纯N_2稀释的单一有机物PTR-MS测量结果表明,在E/N=144 Td条件下,质子化的有机物离子为主要的产物离子峰,没有明显的碎片离子存在。测试了进样系统中催化转化装置的性能,给出了11种挥发性有机物的探测灵敏度和检测下限,讨论了PTR-MS的测量精度。测试结果表明,自行研制的PTR-MS能够实时在线监测《大气污染物综合排放标准》中规定的有机污染物。 Parameters of homemade proton transfer reaction mass spectrometry (PTR-MS) for on-line measurements of volatile organic compounds (VOCs) restricted in "Integrated Emission Standard of Air Pollurants" (GB 16297-1996 ) were optimized and measured. The individual VOCs in pure N2 were measured at E/N = 144 Td by PTR-MS and the results show that the main ions are the protonated compound and few fragmentations exist. The sensitivity and limitation of homemade PTR-MS was obtained to 11 VOCs. The efficiency of the catalytical converter in gas inlet was tested. The precision of the PTR-MS was discussed. The results show that the homemade PTR-MS is a simple and effective tool for real time monitoring the VOCs restricted in "Integrated Emission Standard of Air Pollutants".
出处 《大气与环境光学学报》 CAS 2008年第5期363-368,共6页 Journal of Atmospheric and Environmental Optics
基金 国家自然科学基金(20577049) 中国科学院仪器研制项目(Y2005015) 安徽省优秀青年科技基金(06045098) 合肥物质科学研究院院长基金
关键词 化学电离 质子转移反应质谱 挥发性有机物 在线监测 大气污染物综合排放标准 chemical ionization proton transfer reaction mass spectrometry volatile organic compounds on-line monitoring integrated emission standard of air pollutants
  • 相关文献

参考文献7

  • 1[1]Atkinson R.Atmospheric chemistry of VOCs and NOx[J].Atmospheric Environment,2000,34(12-14):2063-2101. 被引量:1
  • 2[2]Kampa M,Castanas E.Human health effects of air pollution[J].Environmental Pollution,2008,151(2):362-367. 被引量:1
  • 3[3]Lindinger W,Hansel A,Jordan A.Protontransfer-reaction mass spectrometry (PTR-MS):on-line monitoring of volatile organic compounds at pptv levels[J].Chemical Society Reviews,1998,27(5):347-354. 被引量:1
  • 4[4]Hansel A,Jordan A,Holzinger R,et al.Protontransfer reaction mass-spectrometry-online trace gas-analysis at the ppb level[J].International Journal of Mass Spectrometry,1995,150:609-619. 被引量:1
  • 5[6]de Gouw J,Warneke C.Measurements of volatile organic compounds in the earths atmosphere using proton-transfer-reaction mass spectrometry[J].Mass Spectrometry Reviews,2007,26(2):223-257. 被引量:1
  • 6[7]Hayward S,Hewitt C N,Sartin J H,et al.Performance characteristics and applications of a proton transfer reaction-mass spectrometer for measuring volatile organic compounds in ambient air[J].Environmental Science & Technology,2002,36(7):1554-1560. 被引量:1
  • 7[8]Debols D,Giuliani A,Laprevote O.Fragmentation induced in atmospheric pressure photoionization of peptides[J].Journal of Mass Spectrometry,2006,41(12):1554-1560. 被引量:1

同被引文献32

  • 1金顺平,李建权,韩海燕,王鸿梅,储焰南,周士康.质子转移反应质谱在线检测痕量挥发性有机物[J].化学进展,2007,19(6):996-1006. 被引量:20
  • 2Ahn D U, Jo C, Olson D G. Analysis of volatile components and the sensory chaxacteristics of ir- radiated raw pork [J]. Meat Science, 2000, 54(3): 209-215. 被引量:1
  • 3Xu Y, Cheung W, Winder C L, et al. VOC- based metabolic profiling for food spoilage detec- tion with the application to detecting Salmonella typhimurium-contaminated pork [J]. Analytical and Bioanalytical Chemistry, 2010, 397(6): 2439- 2449. 被引量:1
  • 4Mayr D, Margesin R, Klingsbichel E, et al. Rapid detection of meat spoilage by measuring volatile organic compounds by using proton transfer reac- tion mass spectrometry [J]. Applied and Environ- mental Microbiology, 2003, 69(8): 4697-4705. 被引量:1
  • 5Lindinger W, Hansel A, Jordan A. Proton- transfer-reaction mass spectrometry (PTR-MS): on-line monitoring of volatile organic compounds at pptv levels [J]. Chemical Society Reviews, 1998, 27(5): 347-354. 被引量:1
  • 6Shen C Y, Li J Q, Han H Y, et al. Triacetone triperoxide detection using low reduced-field pro- ton transfer reaction mass spectrometer [J]. In- ternational Journal of Mass Spectrometry, 2009, 285(1-2): 100-103. 被引量:1
  • 7Shen C Y, Li J Q, Wang Y J, et al. Discrimination of isomers and isobars by varying the reduced-field across drift tube in proton-transfer-reaction massspectrometry (PTR-MS) [J]. International Jour- nal of Environmental Analytical Chemistry, 2012, 92(3): 289-301. 被引量:1
  • 8Wang Y J, Shen C Y, Li J Q, et al. Thermal des- orption extraction proton transfer reaction mass spectrometer (TDE-PTR-MS) for rapid determi- nation of residual solvent and sterilant in dispos- able medical devices [J]. Journal of Pharmaceuti- cM and Biomedical Analysis, 2011, 55(5): 1213- 1217. 被引量:1
  • 9WangY J, Han H Y, Shen C Y, et al. Control of solvent use in medical devices by proton trans- fer reaction mass spectrometry and ion molecule reaction mass spectrometry [J]. Journal of Phar- maceutical and Biomedical Analysis, 2009, 50(2): 252-256. 被引量:1
  • 10AhnD U, NamK C, Du M, et al. Effect of ir- radiation and packaging conditions after cooking on the formation of cholesterol and lipid oxida- tion products in meats during storage [J]. Meat Science, 2001, 57(4): 413-418. 被引量:1

引证文献3