使用不同检测波长作动态比浊法细菌内毒素检测的比较 被引量:1

Comparison of kinetic turbidimetric limulus assay of bacterial endotoxin with different wavelengths
下载PDF
导出
摘要 目的:探讨不同检测波长对动态比浊法细菌内毒素检测的影响。方法:使用405 nm 及660 nm 的波长动态仪按照2005年版中国药典的方法,对小牛血清(bovine calf sera)及脑蛋白水解物注射液(cerebroprotein hydrolysate injection)样品作动态比浊法细菌内毒素检测。结果:不同波长动态仪测量此2种样品内毒素含量 RSD 小于10%。结论:使用不同波长的动态仪对一般性样品做动态比浊法细菌内毒素检查,检测结果一致。 Objective:To study the effects of different wavelengths on kinetic turbidimetric limulus assay of bacterial endotoxin. Methods : According to ChP 2005 : set the wavelength to 405 nm, and perform separate kinetic turbidimetric limulus assay of bacterial endotoxin both in Bovine Calf Sera and cerebroprotein hydrolysate injection ; set the wavelength to 660 nm and perform the same test. Results:Set the wavelength to 405 nm or 660 nm, all of the RSD of the test concentration on the same wavelength are less than 10%. Conclusion:The test statistics for generic sampies are essentially the same while the wavelengths are set to different.
作者 熊向党
出处 《药物分析杂志》 CAS CSCD 北大核心 2008年第8期1367-1370,共4页 Chinese Journal of Pharmaceutical Analysis
关键词 动态比浊法 细菌内毒素检查 不同波长 kinetic turbidimetric assay bacterial endotoxin test different wavelength