Sysmex XS-800i全自动血细胞分析仪性能评价 被引量:7

Performance Evaluation of Sysmex XS-800i Automated Hematology Analyzer
下载PDF
导出
摘要 目的:评价Sysmex XS-800i全自动血细胞分析仪的性能。方法:对Sysmex XS-800i的准确性、批内、批间精密度、携带污染率、线性、总重复性进行测定,并取60例健康体检标本同时用Sysmex XS-800i和Sysmex 2100测定WBC、RBC、Hb、PLT4项指标并进行比对,并将这60例标本的Sysmex XS-800i分类结果与手工分类结果进行比较。结果:Sysmex XS-800i的准确性、批内、批间精密度、携带污染率、线性、总重复性均在允许范围内(仪器厂家提供),Sysmex XS-800i和Sysmex 2100对WBC、RBC、Hb、PLT的测定结果据统计学分析无显著性差异(P>0.05),Sysmex XS-800i分类结果与手工分类结果进行比较,相关性较好。结论:Sysmex XS-800i是一种较理想的全自动血细胞分析仪。 Objective To evaluate the function of the Sysmex XS-800i automated hematology analyzer. Methods Precision, linearity, carrier rate and repetition were analyzed. Blood samples from 60 healthy people were tested by Sysmex XS-800i and the Sysmex 2100 respectively. Four indexes including WBC,RBC,Hb and PLT from two tests were compared. The results differentiated by Sysmex XS-800i were compared with those by handwork. Results Precision, linearity, carrier rate and repetition studies of Sysmex XS-800i were all in the allowable range (supplied by factory authorities ).The Statistics results of WBC,RBC, Hb and PLT tested by Sysmex XS-800i showed no significant change in comparison to those by Sysmex 2100(P〉0.05). The results differentiated by Sysmex XS-800i showed excellent correlation with those by handwork. Conclusion The performance of the Sysmex XS-800i automated hematology analyzer is satisfied. [Chinese Medical Equipment Journal,2008,29(6) :89-90,93]
出处 《医疗卫生装备》 CAS 2008年第6期89-90,93,共3页 Chinese Medical Equipment Journal
关键词 SYSMEX XS-800I 血细胞分析仪 携带污染率 线性范围 Sysmex XS-800i accuracy precision carrier rate linearity
  • 相关文献

参考文献5

  • 1叶应妩,王毓三主编,中华人民共和国卫生部医政司编..全国临床检验操作规程 第2版[M].南京:东南大学出版社,1997:621.
  • 2Waiters J. Patricia Garrity. Performance evaluation of the Sysmex XE-2100 hematology analyzer[J]. Laboratory hematology,2000,6: 83-92. 被引量:1
  • 3束国防,高茂馗,芦慧霞.XE-2100血液分析仪的应用评价[J].现代医学,2005,33(5):323-325. 被引量:5
  • 4Ruzicka K,Veitl M. The new hematology analyzer Sysmex XE- 2100 [J]. Archives of Pathology & Laboratory Medicine,2001,125 (3) :391. 被引量:1
  • 5Inoue H. Overview of automated hematology analyzer XE-2100[J]. Sysmex J Int, 1999,9:58-64. 被引量:1

二级参考文献5

  • 1LANGFORD K, LUCHTMAN-JONES L, MILLER R, et al. Performance evaluation of the Sysmex XT-2000i automated hematology analyzer [J]. Lab Haematol,2003,9:29-37. 被引量:1
  • 2DARNIGE L, CLUET S, DELAVENNE J, et al. Evaluation of the Beckman Coulter HmX hematology analyser at a general hospital [J]. Ann Biol Clin (Paris),2002,60:47-55. 被引量:1
  • 3GRIMALDI E, SCOPACASA F. Evaluation of the Abbott CELLDYN 4000 hematology analyzer [J]. Am J Clin Pathol,2000,113:497-505. 被引量:1
  • 4PICARD F, GICQUEL C, MARNET L. Preliminary evaluation of the new hematology analyzer COULTER GEN-S in a university hospital [J]. Clin Chem Lab Med, 1999,37:681-686. 被引量:1
  • 5王海亮,束国防,杨润宇,王绍栋.Pentra-60血液细胞分析仪的应用评价[J].铁道医学,2000,28(5):318-319. 被引量:2

共引文献4

同被引文献35

引证文献7

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部