摘要 目的探讨无创心功能监测在冠状动脉旁路移植术后监护中的应用。方法选择2006年1~3月心外ICU收治的29例冠状动脉旁路移植术患者,应用胸腔生物阻抗法心脏血流动力仪和Swan-Ganz导管,同时测定患者的心排出量和肺动脉楔压,并对结果进行对比分析。结果两种方法测定的结果心排出量(CO)分别为(5.03±1.48)L/min和(5.25±1.55)L/min(P>0.05),r=0.82(P<0.001);肺动脉楔压(PCWP)的结果分别为(10.35±2.18)mmHg和(9.86±2.67)mmHg(P>0.05),r=0.71(P<0.05)。两种方法测定的结果具有良好的相关性。结论胸腔生物阻抗法无创心功能监测与Swan-Ganz导管测定结果是一致的,它不仅能准确地测定心排出量和肺动脉楔压,而且还具有无创、连续、简便和廉价的优点。 Objective To evaluate the noninvasive hemodynamic monitoring in patients undergoing coronary artery surgery. Methods Impedance cardiography technology and Swan-Ganz catheter were applied to measure the cardiac output and pulmonary capillary wedge pressure of 29 patients in Cardiac Surgical ICU. Results There was a significant correlation between impedance cardiography technology and Swan-Ganz catheter for cardiac output (5.03±1.48)L/min vs (5.25±1.55)L/min,P〉0.05,r=0.82,P〈0.001, and pulmonary capillary wedge pressure measurement (10.35±2.18)mm Hg vs (9.86±2.67)mm Hg,P〉0.05,r=0.71,P〈0.05). Conclusion There is close agreement between impedance cardiography technology and Swan-Ganz catheter in the measurement of cardiac output and pulmonary capillary wedge pressure in the patients. The impedance cardiography technology is a noninvasive, feasible, handy, continuous and cheap measurement of hemodynamic monitoring.
出处 《中国心血管病研究》 CAS 2008年第2期96-98,共3页 Chinese Journal of Cardiovascular Research
关键词 心电描记术 阻抗 导管插入术 Swan—Ganz 心排血量 肺动脉 肺楔压 Cardiography, impedance Catheterization, Swan-Ganz Cardiac output Pulmonary artery Pulmonary wedge pressure
  • 相关文献

参考文献12

  • 1[1]Chemow B.Pulmonary artery flotation catheters:a statement by the American College of Chest Physicians and the American Thoraeie Society.Chest,1997,111:261-262. 被引量:1
  • 2[2]Wong DH,Tremper KK,Stemmer EA,et al.Noninvasive cardiac output:simultaneous comparison of two different methods with thermodilution.Anesthesiology,1990,72:784-792. 被引量:1
  • 3[3]Tmmm KJ,McCarthy RJ,Spiess BD,et al.Effect of pulmonary artery catheterization on outcome in patients undergoing coronary artery surgery.Anesthesiology,1989,70:199-206. 被引量:1
  • 4[4]Connom AF,Speroff T,Dawson NV,et al.The effectiveness of right heart catheterization in the initial care of critically ill pafients.JAMA,1996,276:889-897. 被引量:1
  • 5[5]Eisenborg PR,Jaffe AS,Schuster DP.Clinical evaluation compared to pulmonary artery catheterization in the hemodynamic assossment of critically ill patients.Crit Care Med,1984,12:549-553. 被引量:1
  • 6卢君强,李春盛,杨佳勇,Jay Zhu,Charles Wo,William C.Shoemaker.无创血流动力学监护系统在高危外科患者复苏治疗中的应用[J].中国危重病急救医学,2003,15(12):730-734. 被引量:24
  • 7[7]Sageman WS,Riffenburgh RH,Spiess BD.Equivalence of bioimpedance and thermodilution in measuring cardiac index after cardiac surgery.J Cardiothorac Vesc Anesth,2002,16:8-14. 被引量:1
  • 8[8]Gerhardt UM,SchoHer C,Bocker D,et al.Non-invasive estimation of cardiac output in critical care patients.Clin Monit Comput,2000,16:263-268. 被引量:1
  • 9[9]Kaukinen S,Koobi T,Bi Y,et al.Cardiac output measurement after coronary artery bypass grafting using bolus themodilution,continuous themodilution,and whole body impedenee eardiography.J Cardiothorac Vesc Anesth.2003,17:199-203. 被引量:1
  • 10[10]Barin E,Haryadi DG,Schookin SI,et al.Evaluation of a thoracic hioimpedance cardiac output monitor during cardiac catheterization.Crit Care Med,2000,28:698-702. 被引量:1