应用EPID分析头颈部肿瘤调强放疗的摆位误差 被引量:21

Analyzing setup error with electron portal images device for head and neck tumors treated with intensity modulated radiation therapy
原文传递
导出
摘要 目的通过对头颈部肿瘤患者群体的射野图像回顾性分析,了解患者群体的摆位误差分布情况,为治疗计划设计设置计划靶体积(PTV)时提供依据。方法通过配准数字重建图像(DRR)和电子射野影像装置(EPID)拍摄的正、侧位验证像的骨性解剖结构,计算平移和旋转误差。结果平移误差左右方向为(1.40±1.27)mm,头脚方向为(1.34±1.37)mm,腹背方向为(1.34±1.30)mm;旋转误差冠状面为(0.791±0.976)°,矢状面为(0.531±0.750)°。结论对于头颈部肿瘤调强放疗(IMRT),临床靶体积(CTV)到PTV的外放边界在左右方向宜为3.7mm,头脚及腹背方向宜为3.6mm。考虑到旋转误差,当靶区比较长时靶区两端外放要更大一些。 Objective To investigate the setup accuracy of the patients with head and neck cancer, and to provide reference data on the margin between the clinical target volume (CTV) and planning target volume(PTN). Methods Electron portal images device (EPID) and digitally reconstructed radiographs (DRR) of patients with head and neck tumors were retrospectively analyzed. The translational and rotational deviations were analyzed by registering and comparing the bony structures of EPID and DRR. Results The translational deviations were ( 1.40 ± 1.27) mm, ( 1.34 ± 1.37) mm and ( 1.34 ± 1.30) mm in medi-lateral, cranio-caudal and anterior-posterior directions, respectively. The rotational deviations were (0.791 ± 0. 976)° and (0. 531 ± 0. 750)° in sagittal plane and coronal plane, respectively. Conclusions For the patients with head and neck tumors undergoing intensity modulated radiation therapy, the margins between the CTV and PTV shoukl be 3.7 mm in lateral direction, and 3.6 mm in both cranio-caudal and anteriorposterior directions. Considering the rotational variations, the margin should be expanded at the end of the target.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2008年第1期39-42,共4页 Chinese Journal of Radiation Oncology
关键词 摆位误差 电子射野影像装置 数字重建图像 调强放射疗法 Setup error Electron portal images device Digitally reconstructed radiographs Intensity modulated radiation therapy
  • 相关文献

参考文献15

  • 1胡逸民 张红志 戴建荣 胡逸民 主编.治疗计划的设计与执行[A].胡逸民,主编.肿瘤放射物理学[C].北京:原子能出版社,1999.463-471. 被引量:8
  • 2Creutzberg CL, Althof VGM, De Hong M, et al. A quality Control Study of the accuracy of patient positioning in irradiation of pelvic fields. Int J Radiat Oncol Phys,1996 ,34 :697-708. 被引量:1
  • 3戴建荣.治疗计划的设计与执行//胡逸民,主编.肿瘤放射物理学.北京:原子能出版社,1999:464. 被引量:2
  • 4Hanley J, Lurrdey MA, Mageras GS, et al. Measurement of patient positioning errors in three-dimensional conformal radiotherapy of the prostate, Int J Radiat Oncol Biol Phys, 1997,37:435-444. 被引量:1
  • 5Hunt MA, Schultheiss TE, Desobry GE, et al. An evalution of setup uncertainties for patients treamed to pelvic sites. Int J Radiat Oncol Biol Phys, 1995,30:227-233. 被引量:1
  • 6Stroom JC, De Boer HC, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability. Int J Radiat Oncol Biol Phys, 1999,43:905-919. 被引量:1
  • 7Remeijer P, Geerlof E, Ploeger L, et al. 3-D portal image analysis in clinical practice : an evaluation of 2-D and 3-D analysis techniques as applied to 30 prostate cancer patients. Int J Oncol Biol Phys, 2000,46 : 1281-1290. 被引量:1
  • 8Fu W, Yang Y, Li X, et al. Dosimetric effects of patient rotational setup errors on prostate IMRT treatments. Phys Med Biol,2006,51: 5321-5331. 被引量:1
  • 9Astreinidou E, Bel A, Raaijmakers CP, et al. Adequate margins for random setup uncertainties in head-and-neck IMRT. Int J Radiat Oncol Biol Phys ,2005 ,61:938-944. 被引量:1
  • 10Langer MP, Papie L, Spirvdovich S, et al. The need for rotational margins in intensity-modulated radiotherapy and a new method for planning target volume design. Int J Radiat Oncol Biol Phys, 2005, 63 : 1592-1603. 被引量:1

共引文献8

同被引文献159

引证文献21

二级引证文献72

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部