便携式静态注射式化学发光分析仪的研制 被引量:3

Development of a Portable Static-Injection Analyzer Based on Chemiluminescence
下载PDF
导出
摘要 针对流动注射化学发光仪存在调试困难和不便于便携等缺点,研制了基于静态注射便携式化学发光分析仪.它由静态注射进样机构、高灵敏度的光电倍增管和单片机PIC18F组成.以Cr3+试验为例,能实现液相化学发光反应的相对发光峰高、相对发光积分面积和发光反应动力学分析.试验表明,用2×10-13g.mL-1的Cr3+标准液作为待测液,它的相对发光峰高和相对发光面积分别为6.0×103,1.6×108. A portable static-injection chemiluminescence instrument for use under field conditions is developed for improving adjustment and operation of flow-injection analyzer. Using a static-injection mechanism as chemiluminescence reaction conduit, we construct a liquid-phase reaction cell and a photo-electric measurement system together with a photomultiplier tube detector, a microcontroller chip of PIC18F and a system software. A direct comparison experiment for measuring Cr^6+ concentration has been carried out to test the performance of the static-injection chemiluminescence instrument in different detection modes such as integral area, light intensity peak and dynamic analysis of chemiluminescence. The relative integral area and light intensity peak are 6.0 × 10^8 , 1.6 × 10^3 respectively, while the sample solution Cr^3+ at the concentration of 2 × 10^-13 g · mL^-1 flows through the reaction cell.
出处 《河南农业大学学报》 CAS CSCD 北大核心 2007年第5期591-594,共4页 Journal of Henan Agricultural University
基金 国家"十五"重大科技专项基金资助(2001BA804A30-08)
关键词 化学发光 光电倍增管 静态注射 便携式 chemiluminescence photomuhiplier tube static-injection portable
  • 相关文献

参考文献7

  • 1LI Q Y,MORRIS K J,DASGUPU P K,et al.Portable flow-injection analyzer with liquid-core waveguide based fluorescence,luminescence,and long path length absorbance detector[J].Analytica Chimica Acta,2003,479:151-165. 被引量:1
  • 2LI B X,ZHANG Z J,ZHAO L X.Flow-injection chemiluminescence detection for studying protein binding for drug with ultrafiltration sampling[J].Analytica Chimica Acta,2002,468:65-70. 被引量:1
  • 3HU Y G,YANG Z Y.A simple chemiluminescence method for determination of chemical oxygen demand values in water[J].Talanta,2004,63:521-526. 被引量:1
  • 4DRAGO K,MATIJA S,JANA K,et al.Chemiluminescence from paper Ⅲ:the effect of superoxide anion and water[J].Polymer Degradation and Stability,2005,88:407-414. 被引量:1
  • 5SHARP J H,BEAUREGARD A Y,BURDIGE D,et al.A direct instrument comparison for measurement of total dissolved nitrogen in seawater[J].Marine Chemistry,2004,84:181-193. 被引量:1
  • 6TORTAJADA L A,FALCO P C,REIG F B.Calibration transfer in chemiluminescence analysis application to chromium determination by luminol-hydrogen peroxide reaction[J].Analytica Chimica Acta,2001,446:385 -392. 被引量:1
  • 7LI J Z,DASGUPTA P K.Measurement of gaseous hydrogen peroxide with a liquid core waveguide chemiluminescence detector[J].Analytica Chimica Acta,2001,442:63-70. 被引量:1