摘要 目的研究一种质控液,它能保证在连续切变毛细管粘度计上测全血粘度获得好的精密度和准确度。方法制备100g/L的聚乙二醇6000做为非牛顿流体(全血)的质控物,检查工作程序和操作技术、校正仪器。结果质控液在三个切变率200s-1,30s-1,3s-1的粘度值分别为4.56±0.12mPa.s,5.18±0.21mPa.s,9.74±0.32mPa.s。批内变异系数(CV)分别为2.50%、3.98%、3.31%;日间CV分别为5.10%、2.70%、4.10%。结论该质控液是一种水溶性液体,与全血对应同属非牛顿流体。为临床血液流变学检验提供了一个实际可用的。 Objective To investigate a special quality control liquid, which ensures high accuracy and precision of whole blood viscosity test with continuous shear capillary viscometer Methods 100 g/L Polyethylene glycol 6000 was prepared and used as 'controls'of non Newtonian liquid and served as a check on procedure,techniques,and instrument calibration.Results The controls indicated three apparent viscosity 4 56±0 116 mPa s,5 18±0 206 mPa s and 9 74±0 323 mPa s under three shear rates at 200 s 1 , 30 s 1 ,and 3 s 1 ,within run cv of 2 5%,3 98%,3 31% and between day CV of 5 1%,2 7%,4 1% respectively Conclusions The control was a water solution ,which provided practical and convenient quality control liquid for routine clinical hemorheological tests
出处 《中华医学检验杂志》 CSCD 1997年第1期41-42,共2页
关键词 血液粘度 质量控制 研究 临床应用 血液学检验 Blood visicosity Quality control