期刊文献+

辐射显色剂量胶片的特性及其在放射医学的应用 被引量:8

Characteristics of radiochromic film and its applications in radiation dosimetry
原文传递
导出
摘要 近十几年来,随着辐射直接显色(RC)剂量胶片灵敏度不断提高,成像技术不断发展,以及商业产品体系的完善,RC剂量胶片在医学领域的应用迅速拓展,成为医用放射剂量测量的有效手段。新类型的RC剂量胶片——EBT,标称测量范围0.1cGy~80kGy,空间分辨率更高,更近组织等效,使小射野剂量测量更准确。这对肿瘤放疗剂量的准确测量、调强放疗剂量分布的精确验证等具有显著优越性。 In recent 10 years, with the increasing of radioehromic (RC) film sensitivity, the development of imaging techniques and improvement of commercial system, RC film expands its application rapidly in medicine area,and becomes an effective method for medical radiation dose measurement. A new type RC film, EBT,is sensitive in dose range from 0. leGy to 80 kGy, and has higher spatial resolution and better tissueequivalence. These characteristics make the dose measurement for small fields more accurate. RC film is superior in exact measurement for tumor radiation dose and accurate verification of dose distribution in intensity modulated radiation therapy.
出处 《国际肿瘤学杂志》 CAS 2007年第1期28-31,共4页 Journal of International Oncology
基金 广东省科技计划基金资助项目(2002C30302)
关键词 X线胶片 放射疗法 辐射剂量 X- Ray film Radiotherapy Radiation dosage
  • 相关文献

参考文献15

  • 1Chiu-Tsao ST,Ho Y,Shankar R,et al.Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies.Med Phys,2005,32(11):3350-3354. 被引量:1
  • 2Cheung T,Butson MJ,Yu PK.Post-irradiation colouration of Gafchromic EBT radiochromic film.Phys Med Biol,2005,50 (20):N281-285. 被引量:1
  • 3Sharma SD,Bianchi C,Conte L,et al.Radiochromic film measurement of anisotropy function for high-dose-rate Ir-192 brachytherapy source.Phys Med Biol,2004,49(17):4065-4072. 被引量:1
  • 4Piermattei A,Fidanzio A,Azario L,et al.A standard dosimetry procedure for 192Ir sources used for endovascular brachytherapy.Phys Med Biol,2002,47 (23):4205-4221. 被引量:1
  • 5Kirisitsa C,Stembergera A,PokrajacaB,et al.Clinical quality assurance for endovascular brachytherapy devices.Radiother Oncol,2004,71(1):91-98. 被引量:1
  • 6Butson MJ,Cheung T,Yu PK,et al.Assessment of large single-fraction,low-energy X-ray dose with radiochromic film.Int J Radiat Oncol Biol Phys,2000,46(4):1071-1075. 被引量:1
  • 7Gamble LM,Farrell TJ,Jones GW,et al.Composite depth dose measurement for total skin electron (TSE) treatments using radiochromic film.Phys Med Biol,2003,48(7):891-898. 被引量:1
  • 8Kubo HD,Araki F.Dosimetry and mechanical accuracy of the first rotating gamma system installed in North America.Med Phys,2002,29(11):2497-2505. 被引量:1
  • 9Daftari I,Castenadas C,Petti PL,et al.An application of GafChromic MD-55 film for 67.5 MeV clinical proton beam dosimetry.Phys Med Biol,1999,44(11):2735-2745. 被引量:1
  • 10Tangboonduangjit P,Metcalfe P,Butson M,et al.Matchline dosimetry in step and shoot IMRT fields:a film study.Phys Med Biol,2004,49 (17):N287 -292. 被引量:1