期刊文献+

迁移管的电场强度对真空紫外电离-离子迁移谱仪性能的影响 被引量:13

Effect of Electric Field of Drift Tube On Performance of A Vacuum Ultraviolet Ionization-Ion Mobility Spectrometer
下载PDF
导出
摘要 离子迁移管是离子迁移谱仪的核心部分,它用来产生均匀的电场,以使不同迁移率的离子进行分离。本研究以丙酮为例,详细研究了本课题组所研制的真空紫外电离源-离子迁移谱仪中迁移管的电场参数对离子的灵敏度和分辨率的影响,发现电压的增大灵敏度增大,但是分辨率存在一个最佳的电压,这些结果可用于迁移谱的优化设计。 The drift tube is the key part of the ion mobility spectrometry. The drift tube is used to produce even electric field, so that different ions can be separated because of different ions mobility. In this study, the acetone was used to illustrate the influence of the electric field strength on the resolving power and sensitivity of vacuum ultraviolet ion mobility spectrometry that is made in our laboratory. It was found that intensity of electric field in the drift tube had a great effect on the ion mobility spectrometry. And there existed a proper electric field strength that could get a more resolving power for ion mobility spectrometry. The conclusions are helpful for design of ion mobility spectrometer.
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 2006年第9期1353-1356,共4页 Chinese Journal of Analytical Chemistry
基金 国家自然科学基金(No.20573111) 863计划(No.2005AA641020)资助课题
关键词 离子迁移谱 迁移管 真空紫外电离 Ion mobility spectrometry, drift tube, vacuum ultraviolet-ionization
  • 相关文献

参考文献12

  • 1Hill H H J,Siems W F,St.Louis R H.Anal.Chem.,1990,62:1201A~1209A. 被引量:1
  • 2许峰,王海龙,关亚风.离子迁移谱研究进展[J].化学进展,2005,17(3):514-522. 被引量:40
  • 3Rubinson K A,Rubinson J F.Contemporary Instrumental Analysis(现代仪器分析).Beijing (北京):Science Press (科学出版社),2003:733~736. 被引量:1
  • 4Eiceman G A,Karpas Z.Ion Mobility Spectrometry,CRC Press,Inc.1994. 被引量:1
  • 5Baim M A,Eatherton R L,Hill H H J.Anal.Chem.,1983,55:1761~1766. 被引量:1
  • 6Leasure C S,Fleischer M E,Anderson G K,Eiceman G A.Anal.Chem.,1986,58:2142~2146. 被引量:1
  • 7Sielemann St,Baumbach J I,Schmidt H,Pilzecker P.Anal.Chim.Acta,2001,431:293~301. 被引量:1
  • 8Vautz W,Sielemann S.Anal.Chim.Acta,2004,513:393~399. 被引量:1
  • 9Vlasak P R,Beussman D J.Rev.Sci.Instrum.,1996,67(1):68~72. 被引量:1
  • 10Siems W F,Wu C,Tarver E E,Hill H H J.Anal.Chem.,1994,66:4195~4201. 被引量:1

二级参考文献86

  • 1Eiceman G A, Karpas Z. Ion Mobility Spectrometry. Boca Raton:CRC Press, 1994.1-228 被引量:1
  • 2Karasek F W. Anal. Chem., 1974, 46: 710A-712A 被引量:1
  • 3Bell S E, Ewing R G, Eiceman G A, Karpas Z. J. Am. Soc., Mass Spectrom., 1994, 5: 177-185 被引量:1
  • 4Fetterolf D D, Donnelly B, Lasswell L D. Proc. SPIE-Int. Soc. Opt. Eng., 1994, 2092: 40-52 被引量:1
  • 5Ewing R G, Miller C J. Field Anal. Chem. Tech., 2001, 15 (5): 215-221 被引量:1
  • 6Eiceman G A, Snyder A P, Blyth D A. Int. J. Environ. Anal. Chem., 1989, 38: 415-425 被引量:1
  • 7Baumbach J I, Eiceman G A. Appl. Spectrosc., 1999, 53: 338A-355A 被引量:1
  • 8Asbury G R, Wu C, Siems W F, Hill H H Jr. Anal. Chim. Acta, 2000, 404: 273-283 被引量:1
  • 9Spangler G E, Collins C I. Anal. Chem., 1975, 47: 403-407 被引量:1
  • 10Purves R W, Guevremont R, Day S, Pipich C W, Maty-jaszczyk M S. Rev. Sci. Instrum., 1998, 69: 4094-4105 被引量:1

共引文献39