期刊文献+

用于硅多条阵列探测器的48路前置电荷灵敏放大器的研制

A 48_channle charge sensitive preamplifier for silicon strip array detector system
下载PDF
导出
摘要 介绍了一套用于硅多条阵列探测器中的高密度多通道电荷灵敏放大器的研制和阵列探测器电子学的发展现状和工作原理,提出了设计原则和设计思路,特别介绍了一个新开发的用于硅多条探测器中的48路电荷灵敏前置放大器的设计和实际测量结果。它具有低成本、高密度的特点,好于1%的能量分辨和良好的长期稳定性。 This paper describes the development of a multi-channle charge sensitive preamplifier used for Silicon Strip Array Detector System,and gives introduction of the current status, working principles and design method for high density muti-channle front-end electrics of Array detector, specially a new devel- oped 48-channel charge sensitive preamplifier for Silicon Strip Array detector System in IMP is introduced which has low-cost, highdensity, 1% energy resoluation and good stability.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2006年第5期568-571,共4页 Nuclear Electronics & Detection Technology
关键词 阵列探测器 多通道 电荷灵敏前置放大器 array detector multi-channel charge sensitive preamplifier
  • 相关文献

参考文献3

  • 1LASSA: a large area silicon strip array for isotopic identification of charged particles[J]. Nucl Instrurn Methods, 2001, A 473 - 302. 被引量:1
  • 2Acosta D, et al. Large CMS Cathode Strip Chambers: design and performance[J]. Nucl Instrurn Methods, 2000, A453 : 182. 被引量:1
  • 3Models 142A, 142B,and 142C Preamlifiers Operating and Service Manual, EG&G ORTEC, 1992, USA 被引量:1