摘要
采用横向加热石墨炉原子吸收光谱法测定轻质碳酸钙中的铅,用磷酸二氢铵和硝酸镁作混合基体改进剂.消除基体干扰.样品铅的测定相对标准偏差为2.24%,回收率为97.8%~98.2%。检出限为3.6pg.
The method of using transversely heated graphite furnace atomic absorption spectrometry (GFAAS) to detect Pb in light calcium carbonate was studied. NH4H2PO4 and Mg(NO3)2 were used as a matrix modifier so as to eliminate the matrix interference. The relative standard deviation of Pb in the sample is 2.24%, the recovery is 97.8-98.2%. and the detection limit of Pb is 3.6 pg.