摘要
目的评价实时三维超声心动图(RT-3DE)估测心室搏出量的可行性和准确性。方法以Swan-Ganz导管热稀释法测定的心室搏出量为标准,应用Philips RT-3DE系统采集10头健康猪心“金字塔形”数据库,结合TomTec4D Cardio-View RT测量软件用心尖长轴8平面法测量左、右室舒张末期和收缩末期容量,计算左、右室搏出量,并与二维超声双平面Si mpson法测值比较。结果所有猪心均获满意的左、右室RT-3DE显示,其形态与直视下猪心心室实际形态相似。其中,RT-3DE8平面法估测左室搏出量[(24.58±3.44)ml]与其右室搏出量[(23.58±3.18)ml]呈高度相关(r=0.93),并与Swan-Ganz导管热稀释法测值[(25.64±3.18)ml]有良好相关性(r=0.85,r=0.92)。经方差分析、SNK均数间两两比较检验,三者间差异无统计学意义(P>0.05)。而二维超声双平面Si mpson法估测左室搏出量[(16.83±4.71)ml]与其右室搏出量[(10.10±2.87)ml]与Swan-Ganz导管热稀释法测值仅呈低度相关(r=0.59,r=0.39),尤以右室搏出量低估程度较明显,且其左、右室搏出量间相关性较低(r=0.76)。经方差分析、SNK检验,三者间差异有统计学意义(P<0.05)。结论RT-3DE是估测心室搏出量的可靠手段。
Objective To evaluate the feasibility and accuracy of real-time three-dimensional echocardiography (RT-3DE) in quantifying ventricular stroke volume in an in vivo experimental study. Methods Ten porcine hearts were examined by RT-3DE and conventional two-dimensional echocardiography (2DE) with Philips Sonos 7500 ultrasound system. The left and right ventricular stoke volume examined by RT-3DE was outlined and calculated by 8-plane method with the 4D Cardio-View RT of TomTec image system, and the results were compared with those of 2DE biplane Simpson-method. The values of ventricular stroke volume measured by Swan-Ganz pulmonary thermodilution method were served as reference values. Results "Full volume" imaging in all porcine hearts was well displayed by RT-3DE. The ventricular shapes displayed by RT-3DE were similar to the anatomic ventricular shapes. From the findings of analysis of variance and Student-Newman-Keuls test, there was no significant difference (P 〉 0.05) between the left ventricular stroke volume with RT-3DE [-(24.58±3.44) ml], the right ventricular stroke volume with RT-3DE [-(23.58±3.18)ml] and those values measured with Swan-Ganz pulmonary thermodilution method [-(25.64±3.18)ml]. Furthermore, there was well correlation between the left ventricular stroke volume and the right ventricular stroke volume with RT-3DE (r = 0.93). They were also well correlated with those values with Swan-Ganz pulmonary thermodilution method (r = 0.85, r = 0.92). However, there was significant difference (P 〈 0.05) between the left ventricular stroke volume with 2DE [-(16.83±4.71)ml], the right ventricular stroke volume with 2DE [(10.10±2.87)ml] and those values with Swan-Ganz pulmonary thermodilution method. Moreover, the right ventricular stroke volume with 2DE was underestimated more than the left ventricular stroke volume with 2DE. There was poor correlation between them (r = 0.76). They were also poorly correlated with those values measured with Swan-G
出处
《中华超声影像学杂志》
CSCD
2006年第6期456-459,共4页
Chinese Journal of Ultrasonography
基金
中国博士后科学基金资助(2004036113)
关键词
超声心动描记术
实时三维
每搏输出量
Echoeardiography, real-time three-dimensional
Stroke volume