摘要
目的探讨STERRAD(r)100S过氧化氢等离子低温灭菌系统在内窥镜消毒中的临床应用效果,并与福尔马林熏蒸灭菌法进行比较。方法2005年1-3月,采用福尔马林熏蒸器械灭菌法完成92台手术104次熏蒸共2706件内窥镜手术器械灭菌。2005年1-3月采用STER-RAD(r)100S过氧化氢等离子低温灭菌系统完成165台手术180个循环共4950件内窥镜手术器械灭菌。对比两种方法的器械损耗率、手术感染率,并发放问卷调查医护人员对两种灭菌方法的满意度。结果STERRAD(r)100S过氧化氢等离子低温灭菌系统在器械损耗率、手术感染率和医护人员满意度指标上均明显优于福尔马林熏蒸灭菌法(P<0.05)。结论STERRAD(r)100S过氧化氢等离子低温灭菌系统具有环保、安全、低温、快速、节省费用和适用范围广的优点,具有极高的经济价值和社会价值。
Objective To evaluate the efficacy of STERRAD (r) 100S (low temperature hydrogen peroxide gas plasma sterilization system) by comparing with formaldehyde steam. Methods From 2005.1 to 2005.3, totally sterilization of 2 760 pieces of operating tools were carried out with formaldehyde steam. From 2005.1 to 2005.3, totally sterilization of 4 950 pieces of operating tools in 180 circles were carried out with STERRAD (r) 100S. The instrument damage rate, operating infection rate and satisfaction rate of medical workers of 2 groups were compared, Results All data were in favor of the efficacy of STERRAD ( r) 100S. Conclusions STERRAD(r) 100Sisasafe, effective, low temperature, and cost- effective sterilization system.
关键词
过氧化氢等离子器
福尔马林
内窥镜
灭菌
Evaluation of the efficacy and usefulness of low temperature hydrogen peroxide gas plasma sterilization system for endoscopies Hydrogen peroxide gas plasma
Formaldehyde
Endoscopes
Sterilization