期刊文献+

心脏起搏器自动阈值夺获功能对使用寿命的影响 被引量:1

下载PDF
导出
摘要 目的对具有自动阈值夺获功能的VVI型心脏起搏器与不具有自动阈值夺获功能的VVI型心脏起搏器在耗电量上的粗略比较。方法对13例具备自动阈值夺获功能的VVI型起搏器和64例不具备自动阈值夺获功能的VVI型起搏器的起搏参数、工作参数等方面的比较分析,以耗电量粗略计算前者能相对延长其使用寿命。结果具备自动阈值夺获功能的VVI型起搏器比不具备自动阈值夺获功能的VVI型起搏器使用寿命延长约87.9%(P<0.01)。结论具备自动阈值夺获功能的VVI型起搏器既能保证起搏的安全,又能大大节省电能,从而明显延长起搏器的使用寿命。
出处 《上海交通大学学报(医学版)》 CAS CSCD 北大核心 2006年第4期433-435,共3页 Journal of Shanghai Jiao tong University:Medical Science
  • 相关文献

参考文献9

  • 1陈灏珠著..心脏导管术的临床应用[M].上海:上海科学技术出版社,1980:492.
  • 2Sermasi S,Marconi M,Libero L,et al.Italian experience with autocapture in conjunction with a membrane lead:pacesetter automatic control of energy and membrane automatic threshold evaluation (Pacemate) study group[J].Pacing Clin Electrophysiol,1996,19(11):1799-1804. 被引量:1
  • 3Schuehert A,Ventura R,Meinertz T.Adjustment of the evoked response sensitivity after hospital discharge in pacemaker patients with automatic ventricular threshold tracking activated[J].Pacing Clin Electrophysiol,2001,24 (2):212-216. 被引量:1
  • 4钱剑安,顾刚,方祖祥.延长起搏器电池寿命的程控和计算方法[J].中国心脏起搏与心电生理杂志,1996,10(3):163-164. 被引量:7
  • 5赵建荣,周志宏,钱剑安.117例低能量起搏患者的长期随访观察[J].中国心脏起搏与心电生理杂志,1998,12(1):26-27. 被引量:3
  • 6Greenberg S,Grossley G,Ellenbogen K,et al.Clinical experience with evoked response and polarization testing in the Microny/Regency SR + Pacemakers:clinical efficacy of autocapture[J].PACE,1997,20(4):1172. 被引量:1
  • 7张琪,章隆泉,周哲慧,刘辉,刘建平,王彬尧.自动夺获起搏器的临床应用[J].上海第二医科大学学报,2000,20(2):132-134. 被引量:2
  • 8沈卫峰主编..实用介入性心脏病学[M].上海:上海科技教育出版社,1997:425.
  • 9钱学贤 杨兴生.现代冠心病监护治疗学[M].北京:人民军医出版社,1994.388. 被引量:4

二级参考文献6

共引文献9

同被引文献23

  • 1CHEN Ruo-han CHEN Ke-ping WANG Fang-zheng HUA Wei ZHANG Shu.Impact of automatic threshold capture on pulse generator longevity[J].Chinese Medical Journal,2006(11):925-929. 被引量:4
  • 2Preston TA, Fletcher RD, Lucchesi BR, et aL Changes in myocardialthreshold : physiologic and pharmacologic factors in patients withimplanted pacemakers[ J]. Am Heart, 1967,74:235. 被引量:1
  • 3耿仁义,朱中林,华伟.实用心脏起撙技术[M].北京:人民军医出版社,2004.39-40. 被引量:1
  • 4Izzet E, Ahmet A, Erten O, et al. Long-term follow-up of pacemakers with an autoeapture pacing system[J ]. Jpn Heart,2002,43: 631. 被引量:1
  • 5Luria D, Gurevitz O, Bar Lev D, et al. Use of automatic threshold tracking function with non-low polarization leads: risk for algorithm malfunction[JJ. PACE,2004,27(4) :453. 被引量:1
  • 6Johannes S, Bernd N,Ewald H, et al. Acute performance evaluation of a new ventricular automatic capture algorithm [J ]. Europace, 2006,8:65. 被引量:1
  • 7Sperzel J, Pitschner HF, Schwarz T, et al. Automatic capture verification in ICD lead systems using intracardiac ventricular evoked response and reduced coupling capacitance [ J ]. Europace,2003,5 : 83. 被引量:1
  • 8Nurnberg J-H, Abdul-Khaliq H, Ewert P, et al. Antibradycardia pacing in patients with congenital heart disease: experience with automatic threshold determination and output regulation (AutocaptureTM [ J ]. Europace ,2003,5 : 199. 被引量:1
  • 9Bauersfeld U, Bernd N,Luciano M, et al. Low-Energy Epicardial Pacing in Children: The Benefit of Autocapture [ J ]. Ann Thorac Surg, 1999,68 : 1 380. 被引量:1
  • 10Cohen MI,Buck K,Tanel RE, et al. Capture management efficacy in children and young adults with endocardial and Unipolar epicardial systems[ J]. Europace, 2004,6:248. 被引量:1

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部