期刊文献+

整体阻抗法与热稀释法测定心输出量的比较 被引量:3

Comparison of Whole-Body Impedance Cardiography with Thermodilution on Measuring Cardiac Output
下载PDF
导出
摘要 目的研究整体阻抗法测定心输出量的临床应用可行性。方法7例多发性外伤患者,经床边应用热稀释法和整体阻抗法同步测定心输出量,每次每种方法均测定3次后取均值,对所得数据进行相关分析和t检验。结果热稀释法与整体阻抗法测定心输出量的相关分析表明二者呈线性相关,相关性良好;热稀释法与整体阻抗法所测心输出量数值接近,无统计学差异(P>0.05)。结论整体阻抗法检测心输出量稳定、可靠且无创伤性,可作为临床重危病人长期连续监测的手段。 Objective To evaluate the reliability of whole-body impedance cardiography( WBIC )and thermodilution(TD) in measuring cardiac output. Methods The cardiac output of 7 patients was measured with WBIC and TD synchronously, each case was examined 3 times with both methods. The 21 pairs of data were then analyzed by correlation and t test. Results The correlation analysis showed that cardiac output measured with TD and WBIC exhibited good linear relationship. The difference in the 21 pairs of data was not significant ( P 〉 0. 05 ). Conclusion Our results suggest that WBIC can be used as a method in monitoring cardic output for patients with critical illness.
出处 《上海交通大学学报(医学版)》 CAS CSCD 北大核心 2006年第3期307-308,共2页 Journal of Shanghai Jiao tong University:Medical Science
关键词 整体阻抗法 热稀释法 心输出量 测定 whole-body impedance cardiography thermodilution cardiac output
  • 相关文献

参考文献4

  • 1Sramek BB.Hemodynamic and pump performance monitoring by electrical bioimpedance[J].ProblRespir Care,1989,2(2):274-290. 被引量:1
  • 2陆瑞昶,王关炎,黄文红.心阻抗图测定左室收缩功能IC指数的临床研究[J].上海第二医科大学学报,1999,19(3):265-267. 被引量:2
  • 3Koobi T,Kaukinen S,Turjanmaa VM,et al.Whole-body impedance cardiography in the measurement of cardiac output[J].Crit Care Med,1997,25(5):779-785. 被引量:1
  • 4Cotter G,Moshkovitz Y,Kaluski E,et al.Accurate,noninvasive continuous monitoring of cardiac output by whole-body electrical bioimpedance[J].Chest,2004,125(4):1431-1440. 被引量:1

二级参考文献2

  • 1陆瑞昶.无损伤心阻抗图测定左室功能[J].中国心血管杂志,1996,1:43-43. 被引量:1
  • 2陆瑞昶,中华心血管病杂志,1996年,1期,43页 被引量:1

共引文献1

同被引文献35

引证文献3

二级引证文献7