摘要 目的应用动态比浊法鲎实验定量测定血液透析滤过液中的细菌内毒素含量。方法对血液透析滤过液中定量添加标准内毒素进行干扰预试验,将血液透析滤过液原液及稀释成5、10、20倍液,定量添加标准内毒素,其回收率分别为 58.17%、106.7%、99.00%和98.79%;从中筛选出最佳的稀释倍数为10倍,进行3个批号血液透析滤过液(040408、 040511、040527)正式干扰试验,血液透析滤过液中定量添加标准内毒素,其回收率分别为76.32%、99.00%和96.24%。结果标准内毒素使用1.00、0.125、0.0156 Eu/ml,将血液透析滤过液稀释成10倍,滤过液中定量添加标准内毒素,回收率均为50%~200%,认为血液透析滤过液对试验无干扰作用,可用于样品检查。结论动态比浊法可以测定血液透析滤过液的细菌内毒素含量。 Objective To determine bacterial endotoxin in the replacement solution of on-line hemodiafiltration (on-line HDF) using kinetic turbidimetric limulus test. Methods and Results Validation test was performed with the replacement solution of on-line HDF in which quantified standard endotoxin was added. The recovery rates ofendotoxin from the replacement solution and its dilutions at 1/5, 1/10, and 1/20 were 58.17%, 106.7%, 99.00% and 98.79%, respectively, suggesting that the optimal dilution was at 1/10. Standard endotoxin was added into the replacement solution of on-line HDF of 3 batches (04040g, 040511,040527), and the recovery rates in their dilution at 1/10 were 76.32%, 99.00% and 96.24%, respectively. The standard endotoxin in the working curve was 1.00, 0.125, and 0.0156 Eu/ml (endotoxin unit/ml), and the dilution at 1/10 of the replacement solution is effective to eliminate the interference in limulus test. Conclusion Kinetic turbidimetric limulus test provide a means to detect endotoxin in the replacement solution of on-line HDF.
出处 《南方医科大学学报》 CAS CSCD 北大核心 2006年第2期242-244,共3页 Journal of Southern Medical University
基金 广东省科技计划项目(2003C33704)~~
关键词 动态比浊法鲎试验 血液透析滤过液 细菌内毒素 干扰试验 kinetic turbidimetric limulus test replacement solution on-line hemodiafiltration bacterial endotoxin inhibition enhancement
  • 相关文献

参考文献5

  • 1[1]Bommer J,Becker KP,Urbaschek R.Potential transfer of endotoxin across high-flux polysulfone membranes[J].J Am Soc Nephrol,1996,7(6):883-8. 被引量:1
  • 2国家药典委员会编..中华人民共和国药典 2005年版 二部[M].北京:化学工业出版社,2005:924.
  • 3[4]U.S.Department of Health and Human Services.Guideline on validation of the LAL test as an end-product endotoxin test for human and animal parenteral drags,biological products and medical devices[C].USDHSS,1987. 被引量:1
  • 4[5]The U.S.Pharmacopeia[M].23rd rev.Easton:Mack Publishing Co.,1995:1969. 被引量:1
  • 5[6]Walhtel RE,Tsuji K.Comparison of limulus amebocyte lysate and the correlation with the United States pharmacopeial pyrogen test[J].Appl Environ Microbiol,1977,33(6):1265. 被引量:1

同被引文献43

  • 1闭闵,李家燕,肖华秀,龚智峰,吴潮清,王浩宇.超纯水对维持性血液透析患者营养不良及贫血的影响[J].广西医学,2004,26(9):1267-1269. 被引量:10
  • 2何东娟,柳宁,陈华光.血液透析患者潜在颅内出血的观察与预防[J].福建中医学院学报,2006,16(2):71-72. 被引量:2
  • 3Jiang Z, Hong Z, Guo W, et al. A symhetie peptide derivod from bactericidal./permeability-increasing protein neutralizes endotoxin in vitro and in vivo[J]. Int lmmunopbarmacol, 2004,4(4) :527-537. 被引量:1
  • 4Rudnick JR, rduino MJ, Land LA, et al .An outbreak of pyrogenic reactions in chronic hemodialysis patients associated with hemodialyzer reuse [J ] .Artif Organs, 1995,9(4) : 89. 被引量:1
  • 5Ledebo I,Ystrand R.Defining the microbiological quality of dialysis fluid[J] .ArtifOrgans, 1999,3(1) :37. 被引量:1
  • 6Kulander L, Nisbeth U, Danielsson BG, et al. Occurrence of endotoxin in dialysis fluid from 39 dialysis units[J ] .J Hosp Infect, 1993,24(1) : 29. 被引量:1
  • 7Arvanitidou M, Spaia S, Askepidis N , et al. Endotoxin concentration in treated water of all hemodialysis units in Greece and inquisition of influencing factors[J] .J Nephrol, 1999,12(1) : 32. 被引量:1
  • 8Bambauer R, Schauer M, Jung WK, et al. Contamination of dialysis water and dialysate.A survey of 30 centers[J ] .Asaio J, 1994,40(4) :1 012. 被引量:1
  • 9山上征二.透析安全基准策定报告.透析会志,1995,28(11):1487-1487. 被引量:1
  • 10林亚丽 颈内静脉置管血透的护理体会[A];中国水利电力医学科学技术学会临床医学专业委员会外科专业、护理专业2007年会论文汇编[C];2007年. 被引量:1