不同起搏方式房室延迟优化的对比研究 被引量:1

Comparison of optimal atrioventricular delay in different pacing modes
下载PDF
导出
摘要 目的观察双腔起搏(DDD)患者不同起搏方式时最佳房室延迟(AVD)的差异。方法对20例置入DDD起搏器的患者经左侧锁骨下静脉插入Swan-Ganz导管,分别测量DDD方式及心房感知心室起搏(VDD)方式下的不同AVD起搏的急性血流动力学效应。结果DDD右心耳起搏优化AVD(149±15ms)比VDD起搏AVD(114±12ms)延长了38±12ms,差异有非常显著性意义(P<0.01)。结论双腔起搏时DDD右心耳起搏可使AVD达到最优。 Objective To investigate optimal atrioventricular delay(AVD) in different pacing modes in patients with dual-chamber (DDD) pacemaker. Methods Acute hemodynamic parameters were measured with Swan-Ganz catheter at varied AVD in DDD and VDD modes in 20 patients. Results Optimal AVD was 149±15ms in DDD fight atrial appendage pacing mode and significantly higher than 114±12ms in VDD pacing mode (P〈0.01). Conclusion DDD right atrial appendage pacing mode makes AVD optimization.
出处 《心电学杂志》 2005年第4期195-197,共3页 Journal of Electrocardiology(China)
基金 陕西省社会发展攻关基金资助项目(编号2004K14-G5)
关键词 双腔起搏 起搏方式 血流动力学 优化房室延迟 Dual-chamber pacing, Pacing mode, Hemodynamics, Optimum atrioventricular delay
  • 相关文献

参考文献8

  • 1Chirfe R,Orega DF,Salazar AL,et al.A method for optimization of left heart atrioventricular interval in patients with dual-cham ber pacemakers. PACE, 1993,16:A185. 被引量:1
  • 2Ishikawa T,Kimura K,Am Nihei T,et al. Relationship between diastolic mitral regurgitation and P-Q intervals on cardiac func tion in patients implanted with DDD pacemakers. PACE,1991,14:1 794. 被引量:1
  • 3Wish M,Fletcher RD,Gottdiener JS, et al.Importance of left atrial timing in the programming of dual-chamber pacemaker. Am J Cardiol,1987,60:566. 被引量:1
  • 4Haskell R J, French WJ.Physiological importance of different atri o-ventricular intervals to improve exercise performance in pa tients with dual chamber pacemakers. Br Heart J, 1998, 61:46. 被引量:1
  • 5Ritter P, Dib JC, Mahaux V, et al. New method for determining the optimal atrioventricular delay in patients paced in DDD mode for complete atrio-ventricular block. PACE,1995,18:855. 被引量:1
  • 6Rey JL, Deschamps PH,Tribouilloy C,et al. Determination of the optimal atrioventricular timing by impedance plethysmography in patients with cardiac pacing correlations with left ventricular filling .Arch Ual Coeur Vaiss,1994,87:737. 被引量:1
  • 7Von Knone GH, Ismer B, Voss W, et al. What rang of programmable AV delays is necessary in antibradycardia DDD stimulation? PACE,1998,21:264. 被引量:1
  • 8Camous JP, Raybaud F,Dolisi C, et al. Interatrial conduction in patients undergoing AV stimulation:effect of increasing right atrial stimulation rate. PACE, 1993, 16:2082. 被引量:1

同被引文献3

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部