常见干扰离子对不同电解质分析仪血清氯测定的影响 被引量:2

Influence of Different Interfering Ions on Determination of Serum Chlorine by Various Ion-Selective Electrodes
下载PDF
导出
摘要 目的:研究常见干扰离子对不同仪器离子选择性电极法血清氯测定的影响。方法:用四种不同的电解质分析仪测定溴化钠、碘化钾、硫化钠以及叠氮钠溶液,并在Cl-均值为100mmol/L的混合血清中分别加入上述四种溶液,观察其具体的干扰情况。结果:所测四种物质均对ISE法测定氯离子产生影响,其中叠氮钠的影响最大;不同的仪器受影响的程度不同,其中以仪器受影响的程度最大。结论:不同的干扰离子对不同的离子选择性电极仪器测血清氯的影响程度不一,我们在工作中应加以注意。 Objective: To study the influence of different interfering ions in determination of serum chlorine by various ion-selective electrodes. Methods:NaBr,KI.Na2S and NaN3 solution were measured by four various electrolyte analyzer, Then joined The four solution were respectively added into a mixed serum with 100 mmol/L of average Cl^- concentration and the disturbing C1- levels were observed, Results:All the four compounds disturbed the determination of serum chlorine by ion selective electrodes (ISE), There were different interferences from the four compounds in which NaN3 exerted the most interference ,and the most extensive influence was occurred in the apparatus Ⅱ , Conclusion :There are different levels of effects from different interfering ions on the determination of serum chlorine by various inhomogeneous ions-selective electrodes. It should be recognized enough in routine of clinical laboratory.
出处 《中国误诊学杂志》 CAS 2005年第13期2416-2417,共2页 Chinese Journal of Misdiagnostics
关键词 氯/血液 离子选择电极 Chlorine/blood Ion-selective electrodes
  • 相关文献

参考文献9

二级参考文献21

共引文献55

同被引文献18

  • 1杨树德.当代临床检验中的固相化学[J].国外医学(临床生物化学与检验学分册),1995,16(3):98-102. 被引量:9
  • 2杨薇,顾健,李玉珍.荧光偏振免疫法在环孢素A血药浓度监测中的质控评价[J].中国医院药学杂志,2005,25(9):820-822. 被引量:7
  • 3Dimeski G,Mollee P,Carter A.Effects of hyperlipidemia on plasma sodium,potassium,and chloride measurements by an indirect lon-selective electrode measuring system[J].Clin Chem,2006,52:155-156. 被引量:1
  • 4Dimeski G,Barnett RJ.Effects of tolal plasma protein concentration on plasma sodium,potassium and chloride measurements by indirect ion selectire electroele measuring system[J].Crit care Resusc,2005,7(1):12-15. 被引量:1
  • 5Hubl W,Wejbora R,Shafti-Keramat I,et al.Enzymatic determination of sodium,potassium,and chloride in abnormal (hemolyzed,icteric,lipemic,paraproteinemic,or uremic) serum samples compared with indirect determination with ion-selective electrodes[J].Clin Chem,1994,40:1528-1531. 被引量:1
  • 6Ono T,Taniguchi J,Mitaumaki H,et al.A new enzymatic assay of chloride in serum[J].Clin Chem,1988,34:552-553. 被引量:1
  • 7Faynor SM,Robinson R. Suitability of plastic collection tubes for cyclosporine measurements[J].Clinical Chemistry,1998,(10):2220-2221. 被引量:1
  • 8Cohen R,Dastane AN. Additive preparation and method of use thereof[P].US,6534016,2003. 被引量:1
  • 9Kessler SB. Apparatus for the separation of blood components[P].US,2578864,1981. 被引量:1
  • 10Anraku H,Shoji Y. Vacuum blood collection tubes[P].US,8565334,1989. 被引量:1