双腔起搏器置入患者最佳房室延迟不同设置方法的临床研究 被引量:4

Clinical Investigation of Different Methods of Determination of the Optimal Atrioventricular Delay in Patients With Implanted Dual-Chamber Pacemakers
下载PDF
导出
摘要 对不同最佳房室延迟(OAVD)设置方法进行比较,探讨OAVD设置的简单、实用的方法.选择36例置入双腔起搏器的患者,设置不同的AVD,于超声心动图下记录左室射血分数(EF)、舒张期二尖瓣返流、A波结束至二尖瓣完全关闭的时间间期(Time 1)、最早出现舒张期二尖瓣返流的AVD(临界AVD),并通过同步心电图测定QT间期(QTI)、QT离散度(QTD)及QRS波群时间(QRSI).将各指标预测的OAVD与OAVD(EF最大时的AVD)进行比较.结果:①心功能正常者心房起搏或心房感知时在AVD为170 ms或140~160 ms时EF达到最大,且与OAVD时EF相比差异均无显著性;心功能不全者在AVD为130~140 ms时EF达到最大,但与OAVD时EF相比差异有显著性.②三种心电图指标预测的OAVD与OAVD比较差异均无显著性,其预测的OAVD在心功能不全患者与OAVD显著相关.③预测OAVD、AVD=200或180 ms时预测的OAVD与OAVD比较差异均无显著性.结论:在实际工作中,可通过下述简单方法设置OAVD: OAVD=200 ms(或180 ms)减去200 ms(或180 ms)时Time 1,以提高随访效率. The objectives of this study is to investigate and compare different methods of determination of the optimal atrioventricular delay (OAVD) and try to find a simple and practical method of identifying OAVD. Thirty-six patients implanted with dual-chamber pacemaker and with ventricular-capture were studied. For each patient, different AVDs were set,and ejection fraction (EF), diastolic mitral regurgitation, the interval between the end of the A wave and complete closure of the mitral valve ( Time 1 ) were measured. At the same time, QT interval, QT dispersion and QRS interval were measured synchronously recorded ECG. Different predictions of optimal AVD were achieved through comparison of all the above-mentioned indexes, and were compared with OAVD(the AVD at which EF was the highest). Results:①In patients with normal caidiac function , EF reached the peak at the AVD 170 ms with atrial-pacing, 140 - 160 ms with atrial-sensing, and there was no significant differences between the EF at the above mentioned AVD and the EF at the OAVD. In patients with insufficient cardiac function, EF reached the peak at the AVD 130 - 140 ms with atrial-pacing, but there was significant difference between the EF at any set AVD and the EF at the OAVD. ②There was no significant difference between the OAVDs predicted by the ECG indexes and the OAVD and there was a significant positive correlation between them in patients with insufficient cardiac function. ③ There was no significant difference between the predicted OAVD, the OAVD predicted when the AVD was set at 200 ms or 180 ms and the OAVD and there was a significant positive correlation between each of them. Conclusions :In our practical work, the OAVD can be calculated by using the follow method: OAVD =200 ms(or 180 ms) subduced the time 1 measuredat the AVD of 200 ms(or 180 ms). Thus, the follow-up can be simplified and more efficient.
出处 《中国心脏起搏与心电生理杂志》 2005年第4期273-276,共4页 Chinese Journal of Cardiac Pacing and Electrophysiology
关键词 心血管病学 双腔心脏起搏器 房室延迟 血流动力学 QT离散度 超声心动图 Cardiology Dual-chamber pacemaker Atrioventricular delay Hemodynamics QT dispersion Echocardiography
  • 相关文献

参考文献8

  • 1Ishikawa T,Sumita S,Kimura K,et al.Prediction of optimal atrioventricular delay in patients with implanted DDD pacemaker[J].PACE,1999,22:1 365. 被引量:1
  • 2朱参战,马奕,崔长琮.心肌细胞电生理和临床心电图进展[J].临床心电学杂志,2002,11(4):232-238. 被引量:9
  • 3Christophe L,David KA.Retiming the failing heart:Principles and current clinical status of cardiac resynchronization[J].J Am Coll Cardiol,2002,39:194. 被引量:1
  • 4Schwaab B,Frohlig G,Alexander C,et al.Influence of right ventricular stimulation site on left ventricular function in atrial synchronous ventricular pacing[J].J Am Coll Cardiol,1999,33:317. 被引量:1
  • 5Ishikawa T, Sugano T, Sumita S, et al.Relationship between atrioventricular delay, QT interval and cardiac function in patients with implanted DDD pacemakers[J].Europace,1999,1:192. 被引量:1
  • 6Lelakowski J, Majewski J, Bednarek J, et al.QT dispersion in DDD and VVI paced patients after myocardial infarction or with left ventricular hypertrophy[J]. Pol Merkuriusz Lek, 2001,11:10. 被引量:1
  • 7Ovsyshcher IE.Toward physiological pacing:Optimization of cardiac hemodynamics by AV delay adjustment[J].PACE,1997,20:861. 被引量:1
  • 8Ishikawa T, Sumita S, Kimura K,et al. Prediction of optimal atrioventricular delay in patients with implanted DDD pacemakers[J].PACE,1999 ,22:1 365. 被引量:1

共引文献8

同被引文献19

引证文献4

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部