摘要
介绍了用于压平眼压计的圆台压平棱镜进行角膜压平直径线性测量原理及定位偏差分析;着重讨论了光通量补偿及角膜压平面积线性测量原理;也介绍了利用计算机图形学设计和制作通量补偿片的方法。
The linear measurement principle of flattened diameter of the ocular cornea and the location deviation of the cone prism which is used in the applanation tonometer are introduced.The principles of luminous flux compensation for the cone prism and linear measurement of flattened area of the ocular cornea are emphatically discussed.The design and making method of the compensating flake of luminous flux using computer graphics is also introduced.
出处
《安徽建筑工业学院学报(自然科学版)》
2004年第6期19-23,共5页
Journal of Anhui Institute of Architecture(Natural Science)
关键词
圆台压平棱镜
定位偏差
光通量补偿
压平面积测量
计算机图形学
cone prism
location deviation
luminous flux compensation
flattened area measurement
computer graphics