期刊文献+

自动洗片机定影液配制及性能测定

The Compounding Method and Charecteristical Test of Fixer Solution for Automatic Processor
原文传递
导出
摘要 作者介绍了自动洗片机定影液的配制方法及性能测定,6种胶片的定透时间及定影效果(密度差)。测定表明:自配定影液在温度30℃以上,冲洗全程150s(定影40s)以上,可满足胶片处理。2年来的实践证明,如此冲洗的胶片质量较好,能达到临床诊断要求。该药液配制方法简单,成本低廉,可替代商品套药在临床上的应用。 This article introduced the prescription and quality control of a fixer solution for au-tomatic processor. The time for fixing and the effectiveness density difference were measured for 6 kinds of films;The results showed that the optimal processing temperature is >30℃, the fotal pro-cessing time is l50s(40s for fixing).Based on 2 years of clinical experience with this fixer solution,the authors concluded that this flxer solution yields films of diagnostic quality, is easily made and of low cost,therefore is a good substitute for the current commercial fixer for automatic processor.
出处 《中华放射学杂志》 CAS CSCD 北大核心 1995年第5期300-302,共3页 Chinese Journal of Radiology
关键词 X线胶片 质量控制 洗片机 定影液 配制 X-ray filmm Quality contro
  • 相关文献

参考文献7

  • 1程相晨,第三军医大学学报,1994年,16卷,126页 被引量:1
  • 2燕树林,中华放射学杂志,1994年,28卷,125页 被引量:1
  • 3曾祥阶,中华放射学杂志,1994年,28卷,77页 被引量:1
  • 4陶可,中华放射学杂志,1992年,26卷,20页 被引量:1
  • 5吴伯卿,中华放射学杂志,1992年,26卷,17页 被引量:1
  • 6李浴中,X线摄影感学化学及显像过程,1992年 被引量:1
  • 7孟代英,X线投照技术,1987年 被引量:1