期刊文献+

应用kin-Com等速装置进行髋关节内旋测试的可靠性 被引量:9

The Reliability for Hip Internal Rotation Measurement of Kin-Com Dynamometer
下载PDF
导出
摘要 本实验应用kin-Com等速装置,对16名男性受试者髋关节内旋功能进行向心收缩-离心收缩交互的测试,以分析其测试的可靠性。通过不同往复运动次数(4次和6次)、实验后重复实验及具体收缩力量的分析等方式测试,结果为不同往复次数所测的力矩值之间无显著性差异,实验后重复实验结果无显著性差异,第1收缩环力量最低,第2次收缩环力量最高,其中第1收缩环向心收缩平均值与其余收缩环的向心收缩平均值有显著性差异,其余则无显著性差异。由此表明,kin-Com等速装置具有较好的稳定性。在运用平均峰值、平均平均值表示测试结果时,在一定范围内选择运动往复次数不致于对其可靠性造成显著的影响。在以向心收缩-离心收缩交互的测试中,牵张反射可能在一定程度上影响力矩值大小。 Hip internal rotation was tested using a concentric eccentric reciprocal testing protocal on the Kin-Com isokinetic dynamometer,In each of two sessions,six and four maximal isokinetic contractions(trials)were elicited from 16 men without isokinetic testing experience.No significant differences occured over sessions and repetitions,but there were significant differences among trials. Significant difference trends occurred in trial 1 and others.Trial 1 was significant lower than others,trial 2 was the highest,but no significant trend was found in trial 2-6.The results demonstrated that a high reliability for isokinetic hip internal rotation strength measurements can be achieved using different repetitions on the Kin-Com device.Significant difference among trials indlcated that torque values associated with strentch reflex when a concentric-eccentric reciprocal testing protocal was used.
出处 《中国运动医学杂志》 CAS CSCD 北大核心 1995年第4期220-222,共3页 Chinese Journal of Sports Medicine
关键词 等速测试 可靠性 髋关节 isokinetic test,reliability
  • 相关文献

参考文献1

  • 1卓大宏主编..中国康复医学[M].北京:华夏出版社,1990:1124.

同被引文献164

引证文献9

二级引证文献166

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部