报告了用紫外光速率法测定血清磷,与比色法对比,r=0.9883,平均回收率为97.5-105%,二份血清CV批内为2.8%、3.0%,批间为4.5%、4.8%,线性范围达3.29mmol/L。研究结果表明:本法具有简单、快速、可靠之优点,适合于自动化分析。
A kinetic for measuring inorganic phosphorus in serum by 240nm wave- length was described. The method compares favoyably with the colorimetric assay(r=0. 9883). avereage recovery is 97. 5-105%. With in run and between-run cvsare 2. 8 to 3.0% and 4.5 to 4.9% respectively in 2 specimens of serum. The stan- dard curve in this method is linear for inorganic phosphorus below 3. 29mmol/L.The mothod is simple. rapid. sensitive. reliable and suitable for Auto assay.