摘要
本文介绍了在0.2mol/LHNO3-0.5mgFe(3+)-0.1mol/L抗坏血酸介质中用电流氧化溶出计时电位法直接测定地质物料中微量铅的新方法。方法检出限0.15μg/g,线性范围2~700ng/ml,Pb的含量水平分别为11.3和58.5μg/g时的精密度(RSD)分别为3.42%和3.08%,回收率95.00%~98.8%。文中对所选测定体系的增敏机理做了研究,考察了40种干扰离子对测定Pb的影响。用本法分析了26个国家一级标准物质中微量Pb,结果与标准值一致。
A new method was developed for determination of trace lead in geological materials by current oxidating stripping chronopotentiometry in the medium of 0.2 mol/L HNO8-0.5mg Fe3+-0.1mol/L ascorbic acid.We have studied the enhancement mechanism of the system and the interferences among more then 40 elements to Pb. Under the optimum conditions selected, the detection limit of Pb is 0.15μg/g in the sample.The dynamic linear working range is 2-700ng/ml.The relative standard deviations (RSD) are 3.42% and 3.08% in the levels of 11.2μg/g and 59.6μg/g, respectively.Twenty-six geochemical reference materials (GSR1-6, GSS1-8, GSD1-12) were determined by the method.The results obtained are good agreements with the certified values.
出处
《冶金分析》
CAS
CSCD
北大核心
1994年第6期1-4,共4页
Metallurgical Analysis
关键词
溶出计时电位法
地质物料
铅
测定
Stripping chronopotentiometry, Geological materials, Lead