期刊文献+

离子电极法与酶法测量钾、钠的比较 被引量:2

Compare Electrode Method with Enzyme Method for K-Na Determination
下载PDF
导出
摘要 目的 对离子电极法和酶法两种测量钾钠的方法作一个评价。方法 通过线性测定、重复 性试验、回收试验、相关性试验、干扰试验对其进行评价。结果 两种方法所检测的项目其线性 范围、回收率均符合要求,重复性良好,批内批间CV%均在允许范围内,相关系数r>0.95,t检 验无显著性差异(p>0.05),酶法的干扰因素大于离子电极法。结论 两种方法可以交替使用,更 加方便了临床应用。 Objective To evaluate the Electrode Method and Enzyme Method for K-Na determination. Method Being evaluated by using linearity test、repetition test、recovery test、correlation test and interference test.Results linearity range and the rate of recovery of two methods is satisfatory. Both of them have a good repeatability and the total imprecision CV% is acceptable, the correlation coefficient r>0.95,t-test has no significant difference(p>0.05).But we should notice that the Enzyme Method for K-Na determination has more interference factor than Electrode Method. Conclusions Both of them can alternate use ,so we can apply them in clinic more conveniently.
出处 《国际医药卫生导报》 2005年第2期91-93,共3页 International Medicine and Health Guidance News
关键词 钾钠测定 离子电极法 酶法 K-Na determination Electrode Method Enzyme Method
  • 相关文献

参考文献5

  • 1叶应妩,王毓三主编,中华人民共和国卫生部医政司编..全国临床检验操作规程 第2版[M].南京:东南大学出版社,1997:621.
  • 2李影林.中华医学会检验全书[M].北京:人民卫生出版社,1996.616-633. 被引量:1
  • 3Berry MN,et al.Clin Chem,1988,34(11):2295 被引量:1
  • 4Kimura S,et al.Clin Chem,1992,38(1):44 被引量:1