期刊文献+

双波长光度法同时测定水中微量铁和铜 被引量:4

SIMULTANEOUS DW-SPECTROPHOTOMETRIC DETERMINATION OF MICROAMOUNTS OF IRON AND COPPER IN DRINKING WATER
下载PDF
导出
摘要 提出了以5 Br PADAP为显色剂,在pH3.0的缓冲溶液中,用双波长光度法同时测定=590.水中微量铁和铜的最佳条件。选择的测定铁和铜的波长对分别为λFe1=503.5nm,λFe2=555.0nm,服从比耳定律的线性范围和检出限分别为0~16μg/25ml和0nm,λCu1=604.0nm,λCu20~24μg/25ml,0.008和0.006μg·ml-1,加标回收率分别为89.2%~105.4%和88.4%~102.0%。通过测定不同的水样,结果满意。 Microamounts of iron and copper in drinking water were determined simultaneously by DW-spectrophotometry using 5-Br-PADAP as chromogenic reagent in a solution buffered at pH 3.0. The conditions and parameters of determination obtained were summarized as follows: for the determination of iron, measuring wavelength: 590.0nm, reference wavelength: 503.5nm; Range of linearity: 0~16μg/25ml, Detection limit: 0.008μg·ml^(-1); Recoveries: 89.2%~105.4%. For the determination of copper: Measuring wavelength: 555.0nm, reference wavelength: 604.0nm; Range of linearity: 0~24μg/25ml, Detection limit: 0.006μg·ml^(-1); Recoveries: 88.4%~102.0%.
出处 《理化检验(化学分册)》 CAS CSCD 北大核心 2004年第12期702-703,705,共3页 Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
关键词 双波长光度法 5-BR-PADAP 水样 DW-Spectrophotometry 5-Br-PADAP Iron Copper Water
  • 相关文献

参考文献2

  • 1Junichiro Miura. Masking Agents in the Spectrophotometric Determination of Metal Ions with 2-(5-Bromo-2-Pyridylazo)-5-Diethylaminophenol and Non-ionic Surfactant[J]. Analyst, 1989,114(10):1323. 被引量:2
  • 2Dralan Menta. Trace Measurement[J]. Laboratory Practice, 1989,3(38):28. 被引量:2

共引文献1

同被引文献940

引证文献4

二级引证文献18

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;