三种黄疸测量仪器对新生儿胆红素测定的比较 被引量:3

A comparison of the difference of three jaundice measure instruments in measuring the bilirubin of neonatal jaundice
下载PDF
导出
摘要 目的:探讨不同仪器对新生儿黄疸胆红素测定的差异.方法:分别应用微量血测胆仪、经皮测胆仪和静脉血自动生化分析仪三种仪器,对30例新生儿进行胆红素测定,并比较其差异.结果:微量血测胆仪与静脉血测黄疸数值无统计学差别;在黄疸程度中等时,经皮测胆仪与静脉血测黄疸数值无明显差异,在黄疸程度较轻或较重时.两者之间出现差异.结论:微量血测胆仪所测胆红素数值较准确,且快捷方便,对新生儿损伤小,适于对黄一疸的监测.具有较深远的临床意义. Objective: To investigate the differences among different bilirubin measure instruments to neonatal jaundice. Methods: Micromethod, transcutaneous bilirubin instruments and vein serological automatic biochemistry analyzer were performed on 30 cases of neonatal jaundice respectively, and the differences were compared. Results: There was no difference of statistical significance between micromethod and serological method; in the cases of middle jaundice, no difference was exist between transcutaneous and vein serological method, but the difference had statistical significance in the cases with slight or serous jaundice. Conclusion: Micromethod is more accurate, convenient and has microlesion to newborns, it has more clinical significance for the measurement to neonatal jaundice.
出处 《现代医学仪器与应用》 2007年第3期54-56,共3页
关键词 胆红素测定 微量血测胆红素 经皮测胆红素 自动生化分析仪 Bilirubin measurement Micromethod bilirubin Transcutaneous bilirubin Automatic biochemistry analyzer