摘要
通过对两台不同型号的露点仪采用不同的气体取样方式进行测试研究,明确了在气体温度和流量相同的情况下,气体压力的变化对露点仪校准结果的影响程度,为露点仪校准提供了适宜的气体取样方式。
Through the test and research of two different types of dew point instruments with different gas sampling methods,the influence of the change of gas pressure on the calibration results of dew point instrument under the same gas temperature and flow is clarified,which provides a suitable gas sampling method for the calibration of dew point instrument.
作者
陈雪霞
陈再举
CHEN Xuexia;CHEN Zaiju(CEPREI,Guangzhou 511370,China)
出处
《电子产品可靠性与环境试验》
2022年第S01期63-66,共4页
Electronic Product Reliability and Environmental Testing
关键词
露点仪
取样方式
压力
dew point meter
sampling method
pressure