期刊文献+

痕量钠离子流动注射离子选择电极分析仪的前期研究 被引量:2

Research on Automatic Analyzer for Determining Trace-level Sodium Based on Flow-injection Ion-selective Electrode Method
下载PDF
导出
摘要 开发了一种用于纯水中痕量Na+的流动注射离子选择性电极快速测定法。以高纯水作载流,实现了在线自动微量碱化、电极自动清洗和完全密封状态下的测定,彻底解决了常规法测量时样品被空气钠离子污染,不能进行准确测定的问题。测定范围为2 0~100μg/L,检出限为1 0μg/L,进样频率为50样/h,一次用样量仅为0 3mL,已成功地用于电厂炉水、蒸汽中痕量Na+的测定。碱化剂用量仅为常规工业钠表的1/50左右。在此基础上开发的新型在线钠表可大大降低耗碱量,提高经济指标,保护环境。 A quick determination method for sodium ion of boiler water and steam at power plants with flow-injection analysis(FIA)and ion selective electrode(ISE)has been expoitured.The basiclize can be accomplished automatically,and the measurement is under sealed circumvent.The electrodes can be rinsed automatically;there is good linearity in the range of 2~100 μg/L.Also,it analyzs very quickly,50 samples/h can be analyzed and the consumption of sample is only 0.3mL/sample.It has been used to determine the trace-level sodium of boiler water and steam at power plants successfully.Alkalization reagent dosage is about 1/50 of general industrial sodium Instrument.The new pattern on-line sodium instrument based on it can reduce alkali dissipation,improve economy index,protect environment.
出处 《仪表技术与传感器》 CSCD 北大核心 2004年第3期24-26,共3页 Instrument Technique and Sensor
关键词 流量注射分析 离子选择电极 钠离子 炉水 蒸汽 发电厂 分析仪 Flow Injecton Analysis Ion Selective Electrode Sodium Ion Hearth Water Boiler Power Plants
  • 相关文献

参考文献10

  • 1LI Y S,GAO X F.Flow injection analysis and application for chemistry analysis.Jilin People's Publisher,2002. 被引量:1
  • 2BULSKA E,WALCERZ M,JEDRAL W,et al.On-line preconcentration of lead and cadium for flame atomic absorption spectromety using a flow-through electrochemical microcell.Anal.Chim.Acta.1997,357:133-140. 被引量:1
  • 3JOHNSON K A,KRIZ D.Development of a three-electrode cell containing a polypyrrol coated auxiliary electrode.Instrumentation Science &Technology,1997,25(1):29-38. 被引量:1
  • 4张利燕.钠离子在线分析技术特点[J].河北电力技术,1999,18(5):31-34. 被引量:4
  • 5李永生 承慰才.流动流射分析[M].北京:北京大学出版社,1986.13209,11-47. 被引量:1
  • 6李永生 承慰才.FIA—TI—721型分光光度分析仪的研制[J].分析仪器,1986,(3):46-46. 被引量:3
  • 7秦玉华,潘洪泽,黄胜全.用于电厂水、汽中微钠离子测定的扩散碱液的研究[J].华北电力技术,2000(5):21-22. 被引量:3
  • 8LI Y S,GAO X F.Flow injection analysis and application for chemistry analysis.Jilin People's Publisher,2002. 被引量:1
  • 9BULSKA E,WALCERZ M,JEDRAL W,et al.On-line preconcentration of lead and cadium for flame atomic absorption spectromety using a flow-through electrochemical microcell.Anal.Chim.Acta.1997,357:133-140. 被引量:1
  • 10JOHNSON K A,KRIZ D.Development of a three-electrode cell containing a polypyrrol coated auxiliary electrode.Instrumentation Science &Technology,1997,25(1):29-38. 被引量:1