A GATE Simulation Study of the Siemens Biograph DUO PET/CT System
1
作者 Dimitrios Nikolopoulos Sofia Kottou +4 位作者 Nikolaos Chatzisavvas Xenophon Argyriou Emannouel Vlamakis Panayiotis Yannakopoulos Anna Louizi 《Open Journal of Radiology》 2013年第2期56-65,共10页
This is a GATE-simulation study of the Siemens Biograph DUO PET/CT system. It reports effects of changes in the thickness of the employed Lutetium Oxyorthosilicate(LSO) detectors. The PET/CT, a human body phantom and ... This is a GATE-simulation study of the Siemens Biograph DUO PET/CT system. It reports effects of changes in the thickness of the employed Lutetium Oxyorthosilicate(LSO) detectors. The PET/CT, a human body phantom and a cylindrical F-18 FDG source were simulated. Validation measurements were conducted. The results indicate that LSO thickness increase degrades spatial resolution, improves relative energy resolution from 9.0% to 11.3% and increases signal-to-noise-ratio from 0.81 to 1.17. Thicker LSO crystals present greater axial sensitivity so as the detection efficiency of PET would be significantly enhanced. 展开更多
关键词 Monte-Carlo GATE pet siemens biograph duo pet/ct