目的探讨高位室间隔植入无导线起搏器的可行性和安全性。方法选取本院参与国内首次无导线起搏器临床研究的患者,比较高位室间隔植入无导线起搏器(A组)与低位室间隔植入者(B组)的有关参数。结果共入选15例患者,均顺利植入Micra起搏器。A...目的探讨高位室间隔植入无导线起搏器的可行性和安全性。方法选取本院参与国内首次无导线起搏器临床研究的患者,比较高位室间隔植入无导线起搏器(A组)与低位室间隔植入者(B组)的有关参数。结果共入选15例患者,均顺利植入Micra起搏器。A组8例,B组7例,A组及B组在植入时的起搏阈值[0.38(0.22)mV vs 0.63(1.00)mV]、R波感知[(10.9±4.7) mV vs (7.3±3.0) mV]、电极阻抗[(906.3±162.4)Ωvs (750.0±173.4)Ω]、T波同步导联数[(7.2±2.7) vs (7.1±2.5)]差异均无显著性(P>0.05)。起搏QRS时限A组有缩短的趋势(140.0 ms vs 179.0 ms,P>0.05),A组起搏QTc短于B组(440.0 ms vs 520.0 ms,P<0.05),A组在手术中的透视时间短于B组[12.86(5.36) min vs 18.46(4.41) min,P<0.05]。两组在出院前、随访1个月时的阈值、感知及阻抗差异均无显著性(P>0.05)。结论高位室间隔植入Micra无导线起搏器安全、可行,起搏的心室除极与复极顺序可能优于低位间隔起搏。展开更多
1 Introduction Over one million cardiac pacemakers are implanted every year worldwide,[11 of which approximately 200,000 are implanted in the United States alone.121 Combined with an aging population and increasing p...1 Introduction Over one million cardiac pacemakers are implanted every year worldwide,[11 of which approximately 200,000 are implanted in the United States alone.121 Combined with an aging population and increasing pacing indications, these numbers are expected to grow. Since the first pacemaker implantation in 1950s, cardiac pacemaker technology has rapidly advanced. Reduction in generator size, increased battery longevity, quality of pacemaker leads, algorithmic and rate responsive programming-all have revolutionized and transformed the implantation and management of transvenous cardiac pacemaker (TV-PPM).展开更多
文摘目的探讨高位室间隔植入无导线起搏器的可行性和安全性。方法选取本院参与国内首次无导线起搏器临床研究的患者,比较高位室间隔植入无导线起搏器(A组)与低位室间隔植入者(B组)的有关参数。结果共入选15例患者,均顺利植入Micra起搏器。A组8例,B组7例,A组及B组在植入时的起搏阈值[0.38(0.22)mV vs 0.63(1.00)mV]、R波感知[(10.9±4.7) mV vs (7.3±3.0) mV]、电极阻抗[(906.3±162.4)Ωvs (750.0±173.4)Ω]、T波同步导联数[(7.2±2.7) vs (7.1±2.5)]差异均无显著性(P>0.05)。起搏QRS时限A组有缩短的趋势(140.0 ms vs 179.0 ms,P>0.05),A组起搏QTc短于B组(440.0 ms vs 520.0 ms,P<0.05),A组在手术中的透视时间短于B组[12.86(5.36) min vs 18.46(4.41) min,P<0.05]。两组在出院前、随访1个月时的阈值、感知及阻抗差异均无显著性(P>0.05)。结论高位室间隔植入Micra无导线起搏器安全、可行,起搏的心室除极与复极顺序可能优于低位间隔起搏。
文摘1 Introduction Over one million cardiac pacemakers are implanted every year worldwide,[11 of which approximately 200,000 are implanted in the United States alone.121 Combined with an aging population and increasing pacing indications, these numbers are expected to grow. Since the first pacemaker implantation in 1950s, cardiac pacemaker technology has rapidly advanced. Reduction in generator size, increased battery longevity, quality of pacemaker leads, algorithmic and rate responsive programming-all have revolutionized and transformed the implantation and management of transvenous cardiac pacemaker (TV-PPM).