A double-sided silicon strip detector(DSSD)with active area of 48 mm x 48 mm and thickness of300μm has been developed. Each side of DSSD consists of48 strips, each with width of 0.9 mm and inter-strip separation of 0...A double-sided silicon strip detector(DSSD)with active area of 48 mm x 48 mm and thickness of300μm has been developed. Each side of DSSD consists of48 strips, each with width of 0.9 mm and inter-strip separation of 0.1 mm. Electrical properties and detection performances including full depletion bias voltage, reverse leakage current, rise time, energy resolution and cross talk have been studied. At a bias of 80 V, leakage current in each strip is less than 15 nA, and rise time for alpha particle at 5157 keV is approximately 15 ns on both sides.Good energy resolutions have been achieved with0.65-0.80% for the junction strips and 0.85-1.00% for the ohmic strips. The cross talk is found to be negligible on both sides. The overall good performance of DSSD indicates its readiness for various nuclear physics experiments.展开更多
A new decay detection system with high detection efficiency and low detection threshold has been developed for charged-particle decay studies, including β-delayed proton, α decay or direct proton emission from proto...A new decay detection system with high detection efficiency and low detection threshold has been developed for charged-particle decay studies, including β-delayed proton, α decay or direct proton emission from proton-rich nuclei. The detection system was tested by using the β-delayed proton emitter ^(24)Si and was commissioned in the decay study of ^(22)Si produced by projectile fragmentation at the First Radioactive Ion Beam Line in Lanzhou. Under a continuous-beam mode, the isotopes of interest were implanted into two double-sided silicon strip detectors, where the subsequent decays were measured and correlated to the preceding implantations by using position and time information. The system allows to measure protons with energies down to about 200 ke V without obvious β background in the proton spectrum. Further application of the detection system can be extended to the measurements of β-delayed proton decay and the direct proton emission of other exotic proton-rich nuclei.展开更多
基金supported by the National Natural Science Foundation of China(Nos.U1432246,U1632136,U1432127,11375268,11635015,and 11475263)the National Basic Research Program of China(No.2013CB834404)
文摘A double-sided silicon strip detector(DSSD)with active area of 48 mm x 48 mm and thickness of300μm has been developed. Each side of DSSD consists of48 strips, each with width of 0.9 mm and inter-strip separation of 0.1 mm. Electrical properties and detection performances including full depletion bias voltage, reverse leakage current, rise time, energy resolution and cross talk have been studied. At a bias of 80 V, leakage current in each strip is less than 15 nA, and rise time for alpha particle at 5157 keV is approximately 15 ns on both sides.Good energy resolutions have been achieved with0.65-0.80% for the junction strips and 0.85-1.00% for the ohmic strips. The cross talk is found to be negligible on both sides. The overall good performance of DSSD indicates its readiness for various nuclear physics experiments.
基金National Basic Research Program of China(2013CB834404)National Natural Science Foundation of China(11375268,11475263,U1432246,U1432127)
文摘A new decay detection system with high detection efficiency and low detection threshold has been developed for charged-particle decay studies, including β-delayed proton, α decay or direct proton emission from proton-rich nuclei. The detection system was tested by using the β-delayed proton emitter ^(24)Si and was commissioned in the decay study of ^(22)Si produced by projectile fragmentation at the First Radioactive Ion Beam Line in Lanzhou. Under a continuous-beam mode, the isotopes of interest were implanted into two double-sided silicon strip detectors, where the subsequent decays were measured and correlated to the preceding implantations by using position and time information. The system allows to measure protons with energies down to about 200 ke V without obvious β background in the proton spectrum. Further application of the detection system can be extended to the measurements of β-delayed proton decay and the direct proton emission of other exotic proton-rich nuclei.