A feasible and convenient method is proposed to suppress higher-harmonics for a varied-line-spacing plane grating monochromator in the soft X-ray region. Related calculations and experiments demonstrate that decreasin...A feasible and convenient method is proposed to suppress higher-harmonics for a varied-line-spacing plane grating monochromator in the soft X-ray region. Related calculations and experiments demonstrate that decreasing the included angle slightly by changing the parameter of the exit arm length can significantly improve light purity. This method is suitable and has been used for experiments of detector calibration in beamline 4B7B at the Beijing Synchrotron Radiation Facility (BSRF).展开更多
基金Supported by National Natural Science Foundation of China(11375227,61204008)
文摘A feasible and convenient method is proposed to suppress higher-harmonics for a varied-line-spacing plane grating monochromator in the soft X-ray region. Related calculations and experiments demonstrate that decreasing the included angle slightly by changing the parameter of the exit arm length can significantly improve light purity. This method is suitable and has been used for experiments of detector calibration in beamline 4B7B at the Beijing Synchrotron Radiation Facility (BSRF).