In trace Li analysis with degenerate four-wave mixing (DFWM) method, acid anions and major metallic elements are dominant interferences in Li-eontaining samples. To better use DFWM technique to analyze trace Li in a...In trace Li analysis with degenerate four-wave mixing (DFWM) method, acid anions and major metallic elements are dominant interferences in Li-eontaining samples. To better use DFWM technique to analyze trace Li in actual samples, we study their effects on Li DFWM signal intensity. It is found that K, Cs, and Ni can enhance the Li DFWM signal, SO4^2, PO^34-, Cl-, and Ca can cause significant suppression, and NO3-, Mg, Ba, Sr, and Na almost have no effects. Finally, we use HaBOa to eliminate the depressive effects of chlorides on Li DFWM signal. The result is also of reference in other trace elements analysis with DFWM.展开更多
基金supported by the National "973" Plan Preliminary Study Special of China(Nos.2012CB723407 and 2010CB434811)the Northwest University Science Foundation(No.13NW12)the Shaanxi Provincial Department of Education Project(No.12JK0599)
文摘In trace Li analysis with degenerate four-wave mixing (DFWM) method, acid anions and major metallic elements are dominant interferences in Li-eontaining samples. To better use DFWM technique to analyze trace Li in actual samples, we study their effects on Li DFWM signal intensity. It is found that K, Cs, and Ni can enhance the Li DFWM signal, SO4^2, PO^34-, Cl-, and Ca can cause significant suppression, and NO3-, Mg, Ba, Sr, and Na almost have no effects. Finally, we use HaBOa to eliminate the depressive effects of chlorides on Li DFWM signal. The result is also of reference in other trace elements analysis with DFWM.