An absolute cryogenic radiometer(ACR) with a detachable optical window was designed and built for high accuracy optical radiant power measurement and photodetector spectral responsivity calibration. The ACR receiver i...An absolute cryogenic radiometer(ACR) with a detachable optical window was designed and built for high accuracy optical radiant power measurement and photodetector spectral responsivity calibration. The ACR receiver is an electroplated pure copper cavity with a 50-μm-thick wall and inner surface coated with a specular black polymer material mixed with highly dispersible carbon nanotubes. The absorptivity of the cavity receivers was evaluated to be ≥0.9999 in the 250 nm–16 μm wavelength range and ≥0.99995 in 500 nm–16 μm. The cavity receiver works at the temperature of ~5.2 K with nanowatt-level noise-equivalent power. The relative standard uncertainty is 0.041% for the measurement of ~100 μW optical radiant power(250 nm–16 μm) and 0.015% for^1 m W(500 nm–16 μm).展开更多
介绍了高准确度光辐射功率校准原理和方法,利用低温辐射计作为主标准器,以陷阱探测器作为传递标准,激光器作为光源,通过激光功率稳定装置,校准了硅陷阱探测器和铟镓砷陷阱探测器的绝对光谱响应度。选取476.1,488,514.7,521,568,632.8,64...介绍了高准确度光辐射功率校准原理和方法,利用低温辐射计作为主标准器,以陷阱探测器作为传递标准,激光器作为光源,通过激光功率稳定装置,校准了硅陷阱探测器和铟镓砷陷阱探测器的绝对光谱响应度。选取476.1,488,514.7,521,568,632.8,647.1,785,852,980,1 064,1 550,nm共12条谱线完成了校准实验,绝对光谱响应度测量不确定度均优于0.05%。通过量子效率模型得出了硅陷阱探测器的绝对光谱响应曲线。利用In Ga As陷阱探测器分立波长点的绝对光谱响应度与相对光谱响应曲线进行了验证分析。结果表明,2种陷阱探测器均可用作传递标准进行高准确度的可见光和近红外光辐射功率校准和量值传递。展开更多
基金supported by the National Key Research and Development Project(No.2016YFF0200301)the National Institute of Metrology Fundamental Research Project(No.AKY1602)
文摘An absolute cryogenic radiometer(ACR) with a detachable optical window was designed and built for high accuracy optical radiant power measurement and photodetector spectral responsivity calibration. The ACR receiver is an electroplated pure copper cavity with a 50-μm-thick wall and inner surface coated with a specular black polymer material mixed with highly dispersible carbon nanotubes. The absorptivity of the cavity receivers was evaluated to be ≥0.9999 in the 250 nm–16 μm wavelength range and ≥0.99995 in 500 nm–16 μm. The cavity receiver works at the temperature of ~5.2 K with nanowatt-level noise-equivalent power. The relative standard uncertainty is 0.041% for the measurement of ~100 μW optical radiant power(250 nm–16 μm) and 0.015% for^1 m W(500 nm–16 μm).
文摘介绍了高准确度光辐射功率校准原理和方法,利用低温辐射计作为主标准器,以陷阱探测器作为传递标准,激光器作为光源,通过激光功率稳定装置,校准了硅陷阱探测器和铟镓砷陷阱探测器的绝对光谱响应度。选取476.1,488,514.7,521,568,632.8,647.1,785,852,980,1 064,1 550,nm共12条谱线完成了校准实验,绝对光谱响应度测量不确定度均优于0.05%。通过量子效率模型得出了硅陷阱探测器的绝对光谱响应曲线。利用In Ga As陷阱探测器分立波长点的绝对光谱响应度与相对光谱响应曲线进行了验证分析。结果表明,2种陷阱探测器均可用作传递标准进行高准确度的可见光和近红外光辐射功率校准和量值传递。