Combining improved injector, gas line and valve-driving models, a gas chromatograph (GC) equipped with Hydrogen Flame Ionization Detector (FID) and Electron Capture Detector (ECD), can measure CH4, CO2, and N2O simult...Combining improved injector, gas line and valve-driving models, a gas chromatograph (GC) equipped with Hydrogen Flame Ionization Detector (FID) and Electron Capture Detector (ECD), can measure CH4, CO2, and N2O simultaneously in an air sample in four minutes. Test results show that the system has high sensitivity, resolution, and precision; the linear response range of the system meets the requirement of flux measurements in situ. The system is suitable for monitoring fluxes of the main greenhouse gases in a short-plant field since it is easy to use, efficacious, and constant and reliable in collecting data.展开更多
文摘Combining improved injector, gas line and valve-driving models, a gas chromatograph (GC) equipped with Hydrogen Flame Ionization Detector (FID) and Electron Capture Detector (ECD), can measure CH4, CO2, and N2O simultaneously in an air sample in four minutes. Test results show that the system has high sensitivity, resolution, and precision; the linear response range of the system meets the requirement of flux measurements in situ. The system is suitable for monitoring fluxes of the main greenhouse gases in a short-plant field since it is easy to use, efficacious, and constant and reliable in collecting data.