期刊文献+

用CLSI EP方案评价总胆红素试剂分析性能 被引量:3

Evaluation of the analytical performance of total bilirubin (TBil) reagents with CLSI method evaluation guidline
下载PDF
导出
摘要 目的通过应用CLSI EP方案分别对国产和进口总胆红素(TBil)试剂的分析性能评价,了解国产总胆红素试剂分析性能能否满足临床需要。方法在同一台全自动生化分析仪上,同时测定试剂准确度、精密度、最低检测限、线性范围、干扰试验、方法比对实验、稳定性观察等7项技术指标。结果两种试剂准确度测定,其相对偏差≤±10%;批间变异系数<4.0%;国产TBil试剂线性范围0.5~865.0μmol/L,进口试剂线性范围0.4~869.0μmol/L;国产TBil试剂的最低检测限为0.5μmol/L,进口试剂的最低检测限为0.4μmol/L;方法比对实验结果Y=1.0069X+0.9366,R2=0.9993;抗干扰能力无明显差异;试剂稳定性相近。结论国产总胆红素试剂与进口试剂的分析性能差异无统计学意义,能满足临床实验室的需要。 Objective Through evaluate the analytical performance of domestic total bilirubin (TBil) reagents and im- posed WAKO total bilirubin (TBil) reagents,undemtanding the performance differences of domestic and imported WAKO TBiL reagents. Methods On the same automatic biochemical analyzer, determinate the reagents' analytical performance at the same time, such as accuracy, precision, minimum detection limit,linear range, anti-interference abiliy, stability, method comparison. Results Determinate accurate of the two reagents, the relative deviasion ≤± 10% ;the within run coefficient of variation ≤4.0% ; The linear range of self-made TBil reagent is 0.5 - 865.0p.mol/L,and the linear range of impored reagent is 0.4 - 869.0panol/ L; The minimum detection limit of domestic TBil reagents was 0.5 μmol/L, of the import reagents was 0.4μmoL/L; the result of the method comparation was Y = 1. 0069X+0. 9366, R2 = 0.9993 ; there was no significant difference with anti-interference ability ; the two reagents had similar stability. Conclusion There was no significant difference of analytical performance with self-made total bilirnbin (TBil) reagents and imported WAKO total bilirubin (TBil) reagents.
作者 刘冰
出处 《四川医学》 CAS 2009年第8期1315-1317,共3页 Sichuan Medical Journal
关键词 CLSI EP方案 钒酸盐氧化法:总胆红素:分析性能评价 CLSI EP vanadate oxidizing method total bilirubin performance evaluation
  • 相关文献

参考文献7

二级参考文献23

  • 1郭健,莫培生,刘宏,白峰,赵海舰,吴锦华.胆红素氧化酶法测定血清结合胆红素[J].中华医学检验杂志,1994,17(5):274-276. 被引量:11
  • 2林其燧 文庆成.临床化学诊断方法大全[M].北京:北京大学出版社,1990.488. 被引量:94
  • 3倪赞明 王琏文 等.五种不同血清总胆红素测定方法筛选和推荐改良J-G法的实验报告[J].中华医学检验杂志,1982,5(1):25-25. 被引量:4
  • 4Nakayma K, Differrences between enzymatic and diazo methods for measuring direct bilirubin[J]. Eur J Clin Chem Clin Biochem, 1995,33: 513-517. 被引量:1
  • 5Kurosaka K, Senba S, Tsubota H, et al. A new enzymatic assay for selectively measuring conjugated bilirubin concentration in serum with use of bilirubin oxidase[J]. Clin Chim Acta, 1998,269:125-136. 被引量:1
  • 6Doumas BT, Yein F, Perry B, et al. Determination of the Sum of bilirubin suggar conjuates in plasma by bilirubin oxidase[J]. Clin Chem, 1999,45:1255-1260. 被引量:1
  • 7Morimoto Y, Ishinar T, Takayma M, et al. Novel assay for measuring serum conjugated bilirubin and its Clinical relevance[J]. J Clinical Laboratary Analysis, 2000, 14: 27-31. 被引量:1
  • 8ビリルビン测定用试药および测定方法[J].日本专利特开平,1997,9-37795. 被引量:1
  • 9Tokuda K, Tanimotok. A new method of measuring bilirubin in serum by Vanadic acid[J]. Jpn J Clin Chem(in Japanes with English abstract), 1993,22:116-122. 被引量:1
  • 10Kojma Ryo , Sasayma Yoshizum, Katayama Koriyama, et al. Measurement of serum bilirubin using sodium nitrite as unique oxidizing agent[J]. Rinsho Kagaku(Nippon Rinsho Kagakkai), 1997,26 (1): 22-30. 被引量:1