期刊文献+

MRI伪影的产生机理及其补偿技术研究 被引量:4

Research of generation mechanism and compensation techniques of MRI artifacts
下载PDF
导出
摘要 分析了MRI伪影的产生机理及采取相应补偿技术并判断是否有效。使用1.0T和1.5T的MRI扫描患者或水模,从改善外界环境、校正软硬件、修改参数或训练患者等角度采取各种补偿技术消除伪影。使伪影得到不同程度的控制,多数图像质量改善明显。通过相应补偿技术,多数伪影可以完全消除,有些伪影仅能减少或位置改变。 Objective To verify the usefulness of the corresponding compensation techniques for MRI artifacts after analyzing their generation mechanism. Methods Different compensation techniques were used to eliminate relative artifacts from the angle of improving background, calibrating software and hardware, editing parameters or instructing patients in scanning the patient or phantom with 1.0TMR and 1.5TMR. Results Artifacts were controlled to different extents and the quality of most MR images were improved. Conclusion Most artifacts can be eliminated completely by the use of corresponding compensation techniques while some others only be reduced a little or shifted.
作者 杨刚 李林
出处 《医疗卫生装备》 CAS 2007年第7期62-63,共2页 Chinese Medical Equipment Journal
关键词 MRI 伪影 机理 补偿技术 MRI artifacts mechanism compensation technique
  • 相关文献

参考文献6

二级参考文献12

  • 1彭振军.医用磁共振成像技术[M].武汉:湖北科学技术出版社,1997.. 被引量:20
  • 2陈星荣 沈天真 等.全声CT和MRI[M].上海:上海医科大学出版社,1994.. 被引量:1
  • 3黄继英 梁星原.磁共振声像原理[M].西安:陕西科学技术出版社,1998.. 被引量:1
  • 4Bullmore E,Brammer M,Rouleau G,et al.Computerized brain tissue classification of magnetic resonance images:a new approach to the problem of partial volume artifact[J].Neuroimage,1995,2(2)∶133-147. 被引量:1
  • 5Nolte UG,Finsterbusch J,Frahm J.Rapid isotropic diffusion mapping without susceptibility artifacts:whole brain studies using diffusion-weighted single-shot STEAM MR imaging[J].Magn Reson Med,2000,44(5)∶731-736. 被引量:1
  • 6Finsterbusch J,Frahm J.Diffusion-weighted single-shot line scan imaging of the human brain[J].Magn Reson Med,1999,42(4)∶772-778. 被引量:1
  • 7Reeder SB,Atalar E,Faranesh AZ,et al.Multi-echo segmented k-space imaging:an optimized hybrid sequence for ultrafast cardiac imaging[J].Magn Reson Med,1999,41(2)∶375-385. 被引量:1
  • 8Halavaara JT,Sepponen RE,Lamminen AE,et al.Spin lock and magnetization transfer MR imaging of local liver lesions[J].Magn Reson Imaging,1998,16(4)∶359-364. 被引量:1
  • 9Yamashita Y,Tang Y,Namimoto T,et al.MR imaging of the liver.comparison between single-shot echo-planar and half-Fourier rapid acquisition with relaxation enhancement sequences[J].Radiology,1998,207(2)∶331-337. 被引量:1
  • 10Chen Q,Siewert B,Bly BM,et al.STAR-HASTE:perfusion imaging without magnetic susceptibility artifact[J].Magn Reson Med,1997,38(3)∶404-408. 被引量:1