伽玛刀3.0T磁共振定位降低皮肤灼伤与感染发生率的方法 被引量:1

Strategy for reduction of skin adustum and infection incidence during 3.0T MR localization before gamma knife radiosurgery
原文传递
导出
摘要 目的控制磁共振射频能量(SAR)影响参数,以降低头部伽玛刀治疗时,在3.0T磁共振定位中局部皮肤的灼伤与感染发生率。方法对200例伽玛刀3.0T磁共振定位患者进行临床随机分组研究,试验组100例患者选择合适扫描序列、调整TR参数等措施以控制射频能量进行磁共振定位,对照组100例患者采用常规方法磁共振定位。结果试验组手术后7例患者有头架固定头钉部的局部轻度灼伤,发生率为7.0%;对照组常规方法定位有28例患者头架固定头钉部的局部轻度灼伤,发生率为28.0%,两组间感染发生率差异有统计学意义(P<0.01)。结论合理调整SAR影响参数,控制射频能量,可以明显控制伽玛刀3.0T磁共振定位中的局部灼伤与感染,使患者缩短住院床日,早日康复出院。 OBJECTIVE To develop the strategy by controlling the related parameter of MR specific absorption rate(SAR) to reduce the skin adustum and infection rate after MR localization. METHODS A total of 200 patients were divided into two groups randomly with 100 patients in each group.The test group receives proper scan sequence,TR during MR localization,while the control group receive normal parameters. RESULTS Totally 7 cases were suffered mild skin adustum where in touch with the iron nails with the incidence of 7.0%.On contrary,28 cases suffered mild adustum in control group with the incidence of 28.0%.The difference in infection rate between two groups was statistically significant. CONCLUSION By properly adjusting the SAR parameters,we can obviously reduce the skin adustum and infection rate of patients,and reduce their days in hospital.
出处 《中华医院感染学杂志》 CAS CSCD 北大核心 2011年第23期4976-4977,共2页 Chinese Journal of Nosocomiology
关键词 射频能量 3.0T磁共振 皮肤灼伤 感染 伽玛刀 Specific absorption rate 3.0TMR Skin adustum Infection Gamma knife
  • 相关文献

参考文献6

  • 1殷蔚伯,余子豪,徐国镇,胡逸民主编..肿瘤放射治疗学 第4版[M].北京:中国协和医科大学出版社,2008:1374.
  • 2倪萍,陈自谦,张鲁闽,马继民.3T磁共振的特点及其质量控制策略[J].中国医疗设备,2009,24(12):121-124. 被引量:20
  • 3Boss A,Graf H, Berger A, et al. Tissue warming andregula- tory responses induced by radio freqiaency energy depositionon a whole-body 3-Tesla magnetic resonance imager[J]. Magn Re-son Imaging, 2007,26 (5) : 1334-1339. 被引量:1
  • 4Stralka JP, Bottomley PA. A prototype RF dosimeter for inde- pendent measurement of the average specific absorption rate (SAR) during MRI[J]. Magn Reson Imaging,2007,26(5) :1296-1302. 被引量:1
  • 5Nadobny J, Szimtenings M, Diehl D,et al. Evaluation of MR- in-duced hot spots for different temporal SAR modes using a time-de-pendent finite difference method with explicit temper- ature gradient treatment[J]. IEEE Trans Biomed Eng, 2007, 54(10) : 1837-1850. 被引量:1
  • 6郑君惠,梁长虹,罗维.高场磁共振射频能量吸收率SAR的研究[J].中国医学影像技术,2008,24(9):1469-1472. 被引量:9

二级参考文献9

共引文献27

同被引文献8

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部