摘要
本研究用瞬态诱发耳声发射(transientevokedotoacousticemission,TEOAE)技术对北京海淀妇幼保健院1998年4~8月出生的新生儿进行听功能筛查,用听觉脑干反应和40Hz听觉相关电位技术作为确诊程序,探讨了TEOAE技术在婴幼儿听功能监测中的作用和可能的应用前景。其结果显示:①1470例新生儿接受了TEOAE听功能筛查,经过双阶段的TEOAE筛查后,总通过率为94.90%;②双阶段TEOAE听功能筛查程序的阳性预测值为73.08%;③接受TEOAE筛查的新生儿,听损伤发病率为9.52%。;其中,高危婴儿室和普通新生儿病房的发病率分别为19.38‰和7.43‰,极重度、重度和轻中度听损伤的发病率分别为1.36‰、2.72‰和4.76‰;④日龄为<1天、1天、2天和≥3天的新生儿的TEOAE筛查通过率分别为62.45%、70.45%、81.58%和84.78%,日龄2天组的通过率高于<1天和1天组(X2=34.430,P=0.000;X2=15.408,P=0.000),而与≥3天组无显著差异(X2=1.393,P=0.238)。
Hearing of 1569 neonates born between April and August of 1998 in Beijing Haidian Health Maternity Hospital were evaluated by transient evoked otoacousticemission (TEOAE) hearing screening. Auditory brainstem responses (ABRs) & 40Hzauditory event related potentials (40Hz-AERPs) were used as diagnostic tests for thosewho had not passed the screening. The results showed as follows.① 1470 (93. 69%)newborns received the screening, 13 of them (19 ears) were diagnosed to have sufferedfrom hearing loss;② The positive expected value of TEOAE screening was 73. 08 %; ③The incidence of hearing loss in neonates was 9. 52%, ④ During the first screening stage(age≤7 day), 2-day old neonatal group showed higher passing less than 1-day & 1-dayold group. Nevertheless,there was no significant difference between 2-day old group andmore than 3-day old group.
出处
《耳鼻咽喉(头颈外科)》
1999年第4期207-211,共5页
Chinese Arch Otolaryngology-Head Neck Surg
基金
北京市自然科学基金!7942009
关键词
瞬态诱发
耳声发射
听力测验
婴儿
新生儿
Transient evoked otoacoustic emissions,TEOAEs
Audiometry,evoked response
Infant,newborn